Basic operation (Continued)

Mark (Red)

Connector Cap

Connector

Camera head

Operation of Micro HD camera

Check system operation according to the following procedures.

1.

Insert the camera connector while aligning it to the red

 

mark as shown on the left figure.

 

When cap is connected to the camera connector:

 

Turn the cap counterclockwise to remove it.

2.

Connect the endoscope adapter to the C mount of camera

CAMERA

OUTPUT LEVEL

WHITE BALANCE

MENU SETTING

POWER

CAMERA

CCU

Camera control unit

DZ-VCA3U

SR-W7MAU (W-VHS)

(Optional)

DT-V1900CG

(Optional)

Isolation transformer AA-V31U

AC 120V

(Optional)

 

head.

3. When operating this system, set the switches on the CCU

panel as follows:

 

1)

OUTPUT LEVEL

: [AUTO]

2)

WHITE BALANCE

: [AUTO]

3)

MENU SETTING

: [OFF]

4.Switch on DZ-VCA3U power.

5.Adjust the white balance.

System operation

Check system operation according to the following procedures.

1.Insert the camera connector while aligning it to the red mark as shown on the left figure.

When cap is connected to the camera connector: Turn the cap counterclockwise to remove it.

2.When the camera is used as a system, connect the power plugs of the camera and other equipments to the isolation transformer AA-V31U.

3.Switch on isolation transformer AA-V31U power switch.

4.Switch on DZ-VCA3U power.

5.Switch on W-VHS power.

Insert W-VHS tape, and make sure that it can record.

6.Switch on the power of monitor.

Check that picture taken by camera can be displayed on the monitor.

7.Operate each device according to its operating procedure.

Note : Consult JVC service personnel or authorized dealer when connecting the system for the first time.

Do not disconnect, or modify the connection by yourself.

18