SERVICING
1.Servicing
Should the product fail to work normally, do not attempt to fix it yourself. Please rely on a qualified service person.
Note:
Before returning the camera head, endoscope adapter, be sure to sterilize the camera head with cable and endoscope.
2.Damage Requiring Service
Unplug this product from the wall outlet and refer servicing to qualified service personnel under the following conditions:
a.When the power supply cord or plug is damaged.
b.If liquid has been spilled, or objects have fallen into the product.
c.If the product has been exposed to rain or water.
d.When the product fails to work normally.
e.If the product has been dropped or damaged in any way.
f.When the product exhibits a distinct change in
3.Operational Life of the Product
Regular maintenance and inspection is essential for ensuring safe and
Inspect and maintain the product before and after use. Also, please be sure to request a qualified service personnel to check the product at least once a year.
STERILIZATION:
The camera head with camera cable and endoscope adapter can be sterilized with ethylene oxide (EO).
EO gas parameters are as follows.
Ethylene oxide gas | : 10 % EO and 90% Inert |
| gases. |
Temperature | : 40 °C to 60° C (104° F to |
| 140° F) |
EO gas intensity | :600mg/L |
Sterilization time | :2 to 4 hours |
Desorption | :Aeration |
Aerator | :50°C and more than 18 |
| hours |
| (In room temperature it is |
| required more than 48 H) |
Notes:
•Always follow the recommended procedure from the manufacturer of your EO sterilization system when sterilizing the
•Never sterilize camera head without connecting endoscope adapter and connector cap.
•Deterioration of camera cable
Due to ethylene oxide, rubber part of camera cable may cause deterioration.
Depending on the frequency of use, however at least once a year, an expert must check the device.
•When connecting the endoscope adapter to the camera head, align it to the screw of C mount on the camera head and turn the endoscope adapter clockwise to tighten it. The O ring of the adapter makes the system waterproof. The connecter of the camera cable should be connected with cap.
Sterilize the system under the above situation.
6