Coding the meter

Code meter before using it for the first time and every time you change to another vial of OneTouch® FastTake® Test Strips.

STEP 1

Enter code mode.

Start with meter turned off. Insert OneTouch® FastTake® Test Strip to turn on meter. All display segments will appear followed by date and time. Code number will appear next.

STEP 2

Match code numbers. Compare code number on meter display with code num- ber on test strip vial. If they match, begin testing. If they do not match, follow Step 3.

STEP 3

Code meter.

Press C button until code numbers match. After appears, OneTouch® FastTake® System is ready for testing.

Display symbols and messages

Blood glucose test result with suggestion to check ketone levels.

Average of last 14 days (64 tests performed).

Blood glucose level higher than 33.3 mmol/L (600 mg/dL).

Blood glucose level lower than 1.1 mmol/L (20 mg/dL).

Temperature too low for testing.

Temperature too high for testing.

Battery power too low for testing. Replace batteries immediately.

Note: Millimoles per litre (mmol/L) is the standard unit of measure in Canada. Select mmol/L whenever you reset the meter.

Problem with meter.*

Meter electronics problem.*

Blood sample applied before appeared on display.*

Test strip damaged, removed during test, or inserted improperly.*

*Retest with a new test strip. If problem persists,call the OneTouch® Customer Care Line at

1 800 663-5521.

ca.OneTouch.www

05266504A AW

10/2003

 

95035 CA Milpitas,

 

2003 .Inc ©LifeScan,

.seconds 15 just in results Accurate 3 STEP

.meter off turn to button M Press .mode memory Exit

3 STEP

.order

.test un d’effectuer avant ®TakeFast ®OneTouch du d'utilisation Guide votre dans lées -détail instructions les Lire: *Important

.test de méthode la et glycémie de lecteur du codage de instructions les l’intérieur à Voir : TEST UN D’EFFECTUER AVANT

.test a doing before Booklet Owner’s ®TakeFast ®OneTouch your in instructions detailed Read *Important:

.procedure test the and meter the coding on instructions for inside See TESTING: BEFORE

.window firmation -con filled completely has blood until strip test of EDGE WHITE TOP to drop blood Hold .strip test against finger your push not Do .strip test of EDGE TOPWHITE to drop blood Touch .sample Apply 2 STEP

in results test 150 last recall to button C pressing Continue .appear will result test recent most next and once ton -but C Press .appears time and date with result test recent most seconds, three After .results test Recall

2 STEP

mémoire*-Aide * e d i u G e c n e r e f Re k c i u Q

douloureux moins Tests / Testing Painful Less

 

.mode memory

 

in are you appears, average

.automatically on turns Meter

day-14 When .button M Press

.off turned meter with Start

.strip test Insert

.mode memory Enter

 

 

1 STEP

 

1 STEP

.meter the into it inserting

 

 

or vial the from it removing

summary

memory meter

when surface its on anywhere

 

 

strip test the touch may you

Test

the Using

hands, dry clean, With Note:

Page 1
Image 1
Lifescan Fast Take manual Step, Summary, Test