42
SCALE OPERATION
CALIBRATION
Scale Calibration Procedure (lbs. or kg) with Remote Display*
1. Temporarily remove the platter and place it in a safe location.
It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights and Measures
statutes and regulations applicable in your area by checking with the appropriate government agency before
placing a newly calibrated unit into service or removing any official seals.
2. If this is a currently installed scanner/Diva scale, there will be either an adhesive label seal or wire sealed cover over the
scale calibration button. New installation units will only have a factory-applied adhesive seal covering the scale calibration
button installed.
To remove the adhesive label seal, peel off the label and discard.
To remove the wire seal, cut the wire and discard. Fully loosen the security screw and lift off the cover.
Place the security screw and cover in a safe location.
Figure 42. Diva Scale security seals
3. Press the calibration button to access the Service Mode. ‘Cal’ then ‘Conf’ will appear on the remote display.
Figure 43. Enter Service Mode
* These procedures are for MS2220 StratosS Scanner/Diva Scale models only.