7.Enter any special calibration frequencies on the extra lines provided and enter the appropriate calibration factors. Delete any standard calibration frequencies that were not used.
8.Click on the Install Probe button. A dialog box will appear asking you if you are sure you want to install the probe.
9.Click on OK. The program will automatically send the calibration information to the meter via the modem and cable. The program will then verify that the probe has been successfully installed in the meter by reading the meter’s memory.
10.The model number and the serial number of the probe just installed will now appear in the table of Installed Probes.
11.To leave the Probe Installation window and return to the main menu click the Close button.
Removing Installed Probes
To remove a specific probe from the meter’s memory...
1.First view an installed probe as described in Viewing Installed Probes.
2.Click on the Remove Probe button. A dialog box will appear asking you to confirm that you want to remove a probe from the meter’s memory.
3.Click OK. The probe will automatically be removed from the meter’s memory.
4.The model number and the serial number of the probe just removed will disappear from the table of Installed Probes.
5.To leave the Probe Installation window and return to the Main Menu Window click the Close button.