© National Instruments Corporation 25 NI 5422 Specifications
Markers
Delay from
Start Trigger to
CH 0 Analog
Output
65 Sample Clock Periods + 110 ns Refer to ts2 at
NI Signal
Generators
Help»Devices»
NI 5422»
NI PXI-5422»
Triggering»
Trigger Timing.
Delay from
Start Trigger to
Digital Data
Output
41 Sample Clock periods + 110 ns
Trigger Exporting
Exported
Trigger
Destinations
A signal used as a trigger can be routed out to any
destination listed in the Destinations specification
of Table9.
Exported
Trigger Delay
65 ns (typical). Refer to ts3 at NI Signal Generators Help»
Devices»NI 5422»NI PXI-5422»Triggering»Trigger
Timing.
Exported
Trigger Pulse
Width
>150 ns. Refer to ts4 at NI Signal Generators Help»
Devices»NI 5422»NI PXI-5422»Triggering»Trigger
Timing.
Table 9.
Specification Valu e Comments
Destinations 1. PFI<0..1> (SMB front panel connectors)
2. PFI<4..5> (DIGITAL DATA & CONTROL front panel
connector)
3. PXI_Trig<0..6> (backplane connector)
Quantity One Marker per Segment.
Quantum Marker position must be placed at an integer multiple of
four samples.
Table 8. (Continued)
Specification Valu e Comments