Chapter 2 Hardware Overview ofthe NI 783xR
© National Instruments Corporation 2-9 NI 783xR User Manual
Figure 2-5. Differential Input Connections for Ground-Referenced Signals
With this connection type, the instrumentation amplifier rejects both the
common-mode noise in the signal and the ground potential difference
between the signal source and the NI783 xR ground, shown as Vcm
in Figure 2-5. In addition, the instrumentation amplifier can reject
common-mode noise pickup in the leads connecting the signal sources to
the device. The instrumentation amplifier can reject common-mode signals
when V+in and V–in (input signals) are both within their specified input
ranges. Refer to Appendix A, Specifications, for more information about
input ranges.
Differential Connections for Nonreferenced or Floating Signal SourcesFigure2 -6 shows how to connect a floating signal source to a channel on
the NI 783xR configured in DIFF input mode.
DIFF Input Mode Selected
–
+
–
+
–
+
–
+
Vcm
Vs
Ground-
Referenced
Signal
Source
Common-
Mode
Noise and
Ground
Potential
I/O Connector
AISENSE
AIGND
VmMeasured
Voltage
Instrumentation
Amplifier
AI+
AI–