Chapter 3 Calibration
NI 783xR User Manual 3-2 ni.com
If you have NI-RIO installed, you can find the internal calibration utility at
Start»All Programs»National Instruments»NI-RIO»device»Calibrate
783xR Device. Device is the NI PXI-783xR or NI PCI-783xR device.
Immediately after internal calibration, the only significant residual
calibration error is gain error due to time and temperature drift of the
onboard voltage reference. You can minimize gain errors by performing an
external calibration. If you are primarily taking relative measurements, then
you can ignore a small amount of gain error and self-calibration is
sufficient.
The Flash memory on the NI 783xR stores the results of an internal
calibration so the CalDACs automatically load with the newly calculated
calibration constants the next time the NI 783xR is powered on.
External CalibrationAn external calibration refers to calibrating your device with a known
external reference rather than relying on the onboard reference. The
NI7 83xR has an onboard calibration reference to ensure the accuracy of
self-calibration. The reference voltage is measured at the factory and stored
in the Flash memory for subsequent internal calibrations. Externally
calibrate the device annually or more often if you use it at extreme
temperatures.
During the external calibration process, the onboard reference value is
re-calculated. This compensates for any time or temperature drift-related
errors in the onboard reference that might have occurred since the last
calibration. You can save the results of the external calibration process to
Flash memory so that the NI 783xR loads the new calibration constants the
next time it is powered on. The device uses the newly measured onboard
reference level for subsequent internal calibrations.
To externally calibrate your device, use an external reference several times
more accurate than the device itself.