BRONCHOSCOPE DISINFECTION GUIDE

EXERABF-160 Series and EVISBF-240 Series Bronchovideoscopes

Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information

WARNING: This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.

HIGH-LEVEL DISINFECTION

1

2

3

4

Suction Cleaning Adapter

Fill basin with disinfectant. Attach a suction cleaning adapter (MAJ-222) to the previously cleaned broncho- scope. Immerse completely in disinfectant solution.

Attach a syringe to the adapter.

While immersed, detach the adapter

Soak all equipment according to

Pull the plunger of the syringe to

from the bronchoscope. Use a clean,

the time and temperature

fill the instrument/suction channels

lint-free cloth to wipe away any bub-

recommended by the disinfectant

and channel cleaning adapter

bles clinging to external surfaces.

manufacturer. Use a timer to

with disinfectant solution.

 

measure the exact contact time.

RINSING

5

 

 

6

 

 

7

To Suction Pump

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Connect the suction cleaning

Disconnect the suction cleaning

Connect suction cleaning adapter

adapter and syringe. Inject air

adapter and immerse the

to bronchoscope and to a tube

to remove disinfectant from the

bronchoscope and adapter in

from a suction pump. Aspirate

instrument/suction channels,

sterile water. Using a sterile,

water through channels for 30

then remove the bronchoscope

lint-free cloth, thoroughly rinse

seconds. Remove from water and

from the disinfectant.

and wipe all surfaces.

aspirate air through channels for

 

 

60 seconds.

8 Suction Cleaning Adapter (MAJ-222)

To Suction Pump

Alcohol

Immerse the bronchoscope's distal end in 70% alcohol and aspirate alcohol for 5 seconds. Remove the distal end from the alcohol and aspirate air for 20 seconds.

DRYING

9

10

TRANSPORTATION

Turn the suction pump off. Hold the control

Dry the bronchoscope and adapter with an

 

section with the openings pointing down and

alcohol-moistened cloth. Make sure that the

Hold all parts of the endoscope when carrying. Do

disconnect the suction cleaning adapter.

equipment is completely dry prior to storage.

not let the ends of the bronchoscope swing freely.

 

 

NOMENCLATURE

 

Single-Use Suction Valve (MAJ-209)

 

Universal Cord

 

Light Guide

Suction Cylinder

 

 

 

 

Remote Switches

 

 

Electrical Connector

Single-Use Biopsy Valve (MAJ-210)

 

 

Instrument/Suction Channel Port

Venting Connector

 

 

Insertion Tube

 

Water-Resistant Cap

Control Section

Boot

 

(MH-553)

 

 

 

UP/DOWN Angulation Control Lever

 

 

Bending Section

Distal Tip

 

 

 

IMPORTANT

Wear appropriate personal protective equipment during reprocessing.

Autoclave all autoclavable accessories.

Store the bronchoscope hanging vertically with all valves and caps removed.

Use a compatible disinfectant and test it for proper concentration as recom- mended by the manufacturer.

Do not use the carrying case to store "patient-ready" endoscopes. Any broncho- scope placed in the carrying case must be reprocessed prior to patient use.

Clean the bronchoscope meticulously prior to disinfection.

© 2004 Olympus America Inc. / Printed in USA 01/04 – PWC-100016

Page 1