BRONCHOSCOPE DISINFECTION GUIDE
EXERA™
Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information
WARNING: This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.
HIGH-LEVEL DISINFECTION
1 | 2 | 3 | 4 |
Suction Cleaning Adapter
Fill basin with disinfectant. Attach a suction cleaning adapter
Attach a syringe to the adapter. | While immersed, detach the adapter | Soak all equipment according to |
Pull the plunger of the syringe to | from the bronchoscope. Use a clean, | the time and temperature |
fill the instrument/suction channels | recommended by the disinfectant | |
and channel cleaning adapter | bles clinging to external surfaces. | manufacturer. Use a timer to |
with disinfectant solution. |
| measure the exact contact time. |
RINSING
5 |
|
| 6 |
|
| 7 | To Suction Pump | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Connect the suction cleaning | Disconnect the suction cleaning | Connect suction cleaning adapter |
adapter and syringe. Inject air | adapter and immerse the | to bronchoscope and to a tube |
to remove disinfectant from the | bronchoscope and adapter in | from a suction pump. Aspirate |
instrument/suction channels, | sterile water. Using a sterile, | water through channels for 30 |
then remove the bronchoscope | seconds. Remove from water and | |
from the disinfectant. | and wipe all surfaces. | aspirate air through channels for |
|
| 60 seconds. |
8 Suction Cleaning Adapter
To Suction Pump
Alcohol
Immerse the bronchoscope's distal end in 70% alcohol and aspirate alcohol for 5 seconds. Remove the distal end from the alcohol and aspirate air for 20 seconds.
DRYING
9 | 10 |
TRANSPORTATION
Turn the suction pump off. Hold the control | Dry the bronchoscope and adapter with an |
|
section with the openings pointing down and | Hold all parts of the endoscope when carrying. Do | |
disconnect the suction cleaning adapter. | equipment is completely dry prior to storage. | not let the ends of the bronchoscope swing freely. |
|
| NOMENCLATURE |
| ||
| Universal Cord |
| Light Guide | ||
Suction Cylinder |
|
|
|
| |
Remote Switches |
|
| Electrical Connector | ||
|
| ||||
Instrument/Suction Channel Port | Venting Connector | ||||
|
| Insertion Tube |
| ||
Control Section | Boot |
| |||
|
|
| |||
UP/DOWN Angulation Control Lever |
|
| Bending Section | Distal Tip | |
|
|
|
IMPORTANT
•Wear appropriate personal protective equipment during reprocessing.
•Autoclave all autoclavable accessories.
•Store the bronchoscope hanging vertically with all valves and caps removed.
•Use a compatible disinfectant and test it for proper concentration as recom- mended by the manufacturer.
•Do not use the carrying case to store
Clean the bronchoscope meticulously prior to disinfection.