M3366/0606 | DPG5000 |
|
|
4. Re-Calibration
This procedure requires a known pressure source of at least ±0.1% accuracy in order to fully utilize the accuracy potential of the DPG5000 (If not available, the gauge can be returned to OMEGA for
Procedure:
A.Ensure the Gauge is at 0 psig (or vacuum if absolute), and adjust the zero as per instructions in step 1.
B.Apply full scale pressure to the pressure port and adjust the span (“S”) pot until the display reads the correct pressure.
C.
D.Repeat steps B and C, until no further adjustment is required.
5. Use Of Peak-Hold Feature (Optional)
After initially turning the gage on (wait until the display stops flashing), a second push of the On/Off button energizes the
6. Analog Output Option (0 To 2 VDC)
The
0 to 2 VDC Table
|
|
| |
|
|
| |
|
|
|
|
| Tip | + Output |
|
|
|
|
|
| Center | – Output |
|
|
|
|
|
| Body | No connection |
|
|
|
|
|
|
|
|
|