10.2.4- JUMPERS CONFIGURATION

 

Function

 

 

 

 

Close Jumper

 

 

 

 

 

 

 

 

220 Ω terminator resistor for the RS422

JP1

(By default opened)

 

switch debounce of 5 Hz for the STOP line

JP2* (By default closed)

 

switch debounce of 5 Hz for the START line

JP3* (By default closed)

 

switch debounce of 5 Hz for the RESET line

JP4* (By default closed)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Jumper

Application

 

 

 

A

Jumper position

B

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

JP5

for STOP line

 

1

KΩ

pull-up *

 

 

1 KΩ

pull-down

JP6

for START line

 

falling edge/low level active*

 

rising edge/high level active

JP7

for STOP line

 

falling edge/low level active*

 

rising edge/high level active

JP8

for RESET line

 

falling edge/low level active*

 

rising edge/high level active

JP9

for START line

 

1

KΩ

pull-up*

 

 

1 KΩ

pull-down

JP10

for RESET line

 

1

KΩ

pull-up*

 

 

1 KΩ

pull-down

 

 

 

 

 

 

 

 

 

 

 

*= Factory set-up

10.2.5- SERIAL DATA OUTPUT CONFIGURATION

 

Data output format

Transmission during

Instrument

with DIP switch 7

 

with DIP switch 7

 

time adjust or in

reference

in position ON

 

in position OFF

 

alarm status

 

(NEWPORT mode)

 

Standard mode

 

 

 

 

 

 

 

 

C2

HHM.MS.S(CR)

 

HH.MM.SS(CR)

--(NUL)--(NUL)--(CR)

C1

HHM.M(CR)

 

HH.MM(CR) *

--(NUL)--(CR)

 

 

 

HH(NUL)MM(CR)**

 

C0

MMS.S(CR)

 

MM.SS(CR)

--(NUL)--(CR)

 

 

 

 

 

*= Data output format only when the colon is lighted.

**= Data output format only when the colon is NOT lighted.

H = Character from "0" to "9" which represent one number of hours.

M = Character from "0" to "9" which represent one number of minutes.

S = Character from "0" to "9" which represent one number of seconds.

Character "." = 02E hex.

Character "-" = 02D hex.

Character "(NUL)" = 0 hex.

Character "(CR)" = 0D hex.