2. Electrode care and application
pH/ORP electrodes are similar to batteries; they age with time and usage. The following information will help maximize electrode life.
2.1 Conditions to Avoid:
•High temperatures, strong acids or caustics will elevate electrochemical reactions and speed electrode aging.
•Coatings on the glass or junction surfaces (i.e. proteins) cause extended response time and inaccurate measurement.
•Never store the electrode tip in deionized (DI) water.
•Never expose electrode to temperatures below
•Never scrape or sand the glass electrode surface.
•Treat glass electrode surfaces with care. The glass is very thin and requires care to prevent accidental breakage.
2.2 Submersible Installation Tips:
•Mount electrodes in a location with ample clearance for removal for periodic cleaning and recalibration. Choose a location that keeps the electrode glass completely submerged at all times.
•Place the electrode tip in pH 4 buffer during system maintenance or storage to avoid dehydration.
•Mount the electrode near tank outlet away from reagent addition areas.
YES
YES
2.3 In-Line Installation Tips:
•pH and ORP electrodes respond best in moving fluids.
•The internal measuring electrode chamber contains a wire within a liquid and a slight amount of air. The electrode must be mounted at least 30° from horizontal to ensure proper sensing. Mounting angles less than 30° will impede performance.
30° | 30° |
| |
NO | NO |
NO
3. pH Electrode Calibration
All pH electrodes are designed to ensure linearity during their lifespan. The following sections define proper electrode operation.
3.1 Offset (STD)
Electrode offsets occur due to:
•Clogged reference junction
•Aged or contaminated reference solution/wire
Check offsets in a pH 7 buffer @ 25 °C. The theoretical output is 0 mV. Any deviation from 0 mV is the pH electrode offset.
pH Electrode Offset pH 7 buffer @ 25°C | Theoretical: | 7.0 pH (0.0 mV) |
Theoretical mV Values @ 25°C
pH | mV |
2 | + 295 mV |
3 | + 236 mV |
4 | +177 mV |
5 | +118 mV |
6 | + 59 mV |
7 | 0 mV |
8 | - 59 mV |
9 | - 118 mV |
10 | - 177 mV |
11 | - 236 mV |
12 | - 295 mV |
New: | pH 7 | ± 0.25 pH (±15 mV) |
Reliable: | pH 7 | ± 0.85 pH (± 50 mV) |
Electrode offsets greater than 0.85 pH (50 mV) indicate the electrode requires cleaning or replacement. See section 5: Maintenance and | |||||||||||||
cleaning. |
|
|
|
|
|
|
|
|
|
|
|
| |
| °C |
|
|
|
| pH Error |
|
|
|
| |||
3.2 Slope (SLP) |
|
|
|
|
|
|
|
|
|
|
|
| |
| 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | ||
|
| ||||||||||||
Electrode slope is the number of mV per pH unit. At 25°C the | 15 | 0.15 | 0.12 | 0.09 | 0.06 | 0.03 | 0 | 0.03 | 0.06 | 0.09 | 0.12 | 0.15 | |
theoretical slope is 59.16 mV per pH. Temperature has an | |||||||||||||
25 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | ||
appreciable affect on electrode slope. Reliable instrumentation | |||||||||||||
includes temperature compensation. The graph below illustrates | 35 | 0.15 | 0.12 | 0.09 | 0.06 | 0.03 | 0 | 0.03 | 0.06 | 0.09 | 0.12 | 0.15 | |
potential pH error when a temperature compensated instrument | 45 | 0.3 | 0.24 | 0.18 | 0.12 | 0.06 | 0 | 0.06 | 0.12 | 0.18 | 0.24 | 0.3 | |
is not used. | 55 | 0.45 | 0.36 | 0.27 | 0.18 | 0.09 | 0 | 0.09 | 0.18 | 0.27 | 0.36 | 0.45 | |
| |||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
Recommendations:
•Calibrate temperature before calibrating the standard and slope.
•The mV offset will track across the entire pH range. The slope is usually not affected by offset changes. (i.e. pH 7= +10 mV, pH 4= +187 mV); slope = 59 mV
•Coatings on the glass may affect sensor slopes. See section 5 maintenance and cleaning.
•A constant output near 0 mV in all buffer solutions indicates a shorted electrode that must be replaced.
3.3 Response Time/Stability
Response time and stability are affected by the condition of the pH electrode's glass surface (ORP electrode - Platinum surface), reference junction, and reference solution. Restoration to acceptable levels can often be accomplished by cleaning the electrode's glass surface (ORP electrode - Platinum surface) and reference junction.
Electrode mV values should remain stable ±3 mV. Conditions that may cause fluctuations are:
1.Electrode coating
2.Ground fault: If proper operation is observed in the beaker, but is unstable in the application, a ground fault probably exists.
•Using instrumentation with isolated inputs and outputs may restore stable operation.
•Solution grounding may also restore stable operation.
page 4 | OMEGA PHE/ORE Series pH/ORP Electrode Instructions |