4. ORP Electrode Calibration

All ORP electrodes are designed to ensure linearity during their lifespan. The following sections define proper electrode operation.

4.1 Offset (STD)

Electrode offsets occur due to:

Clogged reference junction

Aged or contaminated reference solution/wire

Offsets are easily checked in pH 7 buffer saturated with quinhydrone @ 25 °C; since the theoretical output is +87 mV. Any deviation from +87 mV is the ORP electrode offset (i.e. +90 mV). Quinhydrone is the oxidizer measured by the ORP electrode and is required for calibration. To guarantee buffer saturation, mix 1/8g quinhydrone per 50 mL of pH buffer.

ORP Electrode Offset:

Solution: pH 7 or pH 4 buffer saturated with Quinhydrone @ 25 °C

 

pH 7 + Quinhydrone

pH 4 + Quinhydrone

Theoretical mV:

+ 87 mV

+ 264 mV

New:

87 ± 15 mV

+264 ± 15 mV

Reliable:

87 ± 50 mV

+264 ± 50 mV

Electrode offsets greater than ±50 mV indicate the electrode requires cleaning or replacement, see section 5.2.

4.2 Slope (SLP)

ORP slope errors are generally caused by contamination of the platinum electrode surface. Cleaning the electrode surface will usually restore proper values, response time, and stability.

 

Common ORP Values

 

Reaction

 

mV

Cr Cr2+ + 2e-

- 913

Fe Fe2+ + 2e-

- 440

Cr2+ Cr3+ + e-

- 407

4OH- O2 + 2H2O + 4e

- 401

2I- I2 + 2e-

- 400

Ti2+ Ti3+ + e-

- 370

Ni Ni2+ + 2e-

- 250

Pb Pb2+ + 2e-

- 126

Fe Fe3+ + 3e-

- 37

H 2H+

+ 2e-

0

2

 

 

Fe2+ Fe3+ + e-

+ 771

Ag Ag+ + e-

+ 799

Pb Pb4+ + 4e-

+ 800

3Br- Br3- + e-

+ 1060

2Br- Br2 + 2e-

+ 1066

ClO - ClO + e-

+ 1160

2

2

 

Pt Pt2+ + 2e-

+ 1188

Ag Ag2+ + 2e-

+ 1369

 

 

 

Many systems require both pH and ORP calibration. To conserve calibration reference solutions, use pH 7 and 4 buffers for pH calibration first. ORP calibration can be performed with the same buffers by adding quinhydrone.

5. Maintenance and Cleaning

5.1 Maintenance

Variables can affect long term pH or ORP electrode life. For this reason, a maintenance log is recommended for trend analysis. When storing boxed sensors, lay the sensor flat to maximize hydration of the glass surface. Keep the glass surface wet at all times. Soak the sensor tip in pH 4.0 buffer during system maintenance intervals. In-line applications should be plumbed with a depression (trap) so liquid is maintained around the sensor tip. If the sensor dehydrates, soak the sensor tip in pH 4 buffer for 24 to 48 hours, then visually inspect the electrode for surface cracks, swelling, or discoloration. Severely dehydrated electrodes cannot be restored to normal operation.

OMEGA PHE/ORE Series pH/ORP Electrode Instructions

page 5

Page 5
Image 5
Omega PHE/ORE Series manual ORP Electrode Calibration, Maintenance and Cleaning, PH 7 + Quinhydrone PH 4 + Quinhydrone