4Calibration Instructions

Table 4-1. Calibration Values for the TX92

 

Temperature

 

mV INPUT

 

mV INPUT

 

Input Range

Model

REF 32°F

Model

REF 32°F

 

Zero/Span

TX92

Zero/Span

TX92

Zero/Span

 

0/1200°F

-J2

-0.855/14.906

-K2

-0.692/13.819

 

0/1300°F

-J3

-0.885/17.947

-K3

-0.692/16.092

 

0/1500°F

-J4

-0.885/14.108

-K4

-0.692/10.560

 

0/1750°F

-J5

-0.885/20.406

-K5

-0.692/15.295

 

0/1000°F

-J6

-0.885/29.515

-K6

-0.692/22.251

 

0/1200°F

-T2

-0.674/13.967

-E2

-1.026/15.869

 

0/1300°F

-T3

-0.674/16.647

-E3

-1.026/19.708

 

0/1500°F

-T4

-0.674/12.572

-E4

-1.026/17.942

23

0/1750°F

--

--0.674/12.572

-E5

-1.026/26.858

0/1000°F

--

--0.674/12.572

-E6

-1.026/40.056

Page 28
Image 28
Omega TX91 manual Calibration Values for the TX92