For digital probe matching within the instrument (version 2.1 only) additional commands are used as follows:
CORRECTIONS A | (turns OFF B corrections) | Turns A on |
B | (turns OFF A corrections) | Turns B on |
AB | (turns ON BOTH A & B corrections) | |
OFF | (turns OFF BOTH A & B corrections) |
Whatever is set by this command is retained during
LA | Lists Ch A corrections | |
LB | Lists Ch B corrections | |
CA | Enter Ch A corrections | |
CB | Enter Ch B corrections | |
Query | Says “Not enabled”, “A”, “B”, “AB” | |
XX* | Disables corrections | |
ENABLE* | Enables corrections | |
IC | Initialise corrections (i.e.Erase) – useful after installing software | |
*For these commands to be actioned the SPANs key must be pressed as the | ||
command is typed. |
| |
Corrections entered as Follows |
| |
|
|
|
Type |
| TM Sends |
CA <CR> |
| How Many Points |
2 <CR> |
| Entry=/0 Ohms |
100,000 <CR> | Deg C? | |
0.000 <CR> |
| 0100.0000 Ohms 0000.0000 Deg C ok? |
Y <CR> |
| Entry=/1 Ohms |
138.510 <GR> | Deg C? | |
100.000 <CR> | 0138.5100 Ohms 0100.0000 Deg C ok? | |
Y <CR> |
| Done |
|
|
|
This would repeat for more points.
Pressing N at the ok? Prompt will allow
Once entered, type LA OR LB TO SEE/CONFIRM CONTENTS.
Rules of entry.
1)Ohmic values must always increment from point to point
2)All values must be entered to THREE decimal places