Omron HBP-1300 instruction manual Environmental Conditions, Non-Invasive Blood Pressure Nibp

Models: HBP-1300

1 38
Download 38 pages 5.6 Kb
Page 33
Image 33

Environmental Conditions

Operational temperature and

Temperature range: 10 to 40°C (50 to 104°F)

humidity

Humidity range: 30 to 85%RH (not condensed)

 

Atmospheric pressure: 700 to 1060hPa

 

 

Storage and transportation

Temperature range: -20 to 60°C (-4 to 140°F)

 

Humidity range: 10 to 95%RH (not condensed)

 

Atmospheric pressure: 500 to 1060hPa

 

 

Non-Invasive Blood Pressure (NIBP)

Measurement technology

Oscillometric

Measurement method

Dynamic Linear Deflation method

 

 

Pressure display range

0 to 300 mmHg

 

 

Pressure display accuracy

Within ±3 mmHg

 

 

NIBP measurement range

SYS 60 to 250 mmHg

 

DIA 40 to 200 mmHg

 

Pulse rate 40 to 200 /min

 

 

NIBP accuracy*

Maximum mean error within ±5mmHg

 

Maximum standard deviation within 8mmHg

 

 

Pulse rate accuracy

Within ±5 % of reading

 

 

Reference standard:

EN1060-1:1995+A2:2009

 

EN1060-3:1997+A2:2009

 

ISO81060-1:2007

 

 

*Comparison with auscultation method performed by a trained professional. DIA determined by the auscultation method is “K5”.

Note: Specifications may be changed without prior notice.

EN

This blood pressure monitor fulfils the requirements of the EC directive 93/42/EEC (Medical Device Directive). It also conforms to the European standard EN 1060, Non-invasive Sphygmomanometers Part 1: General Requirements and Part 3: Additional Requirements for Electromechanical Blood Pressure Measuring Systems.