Manufacturer’s Declaration

The HBP-1300 is intended for use in the electromagnetic environment specified below.

The customer or the user of the HBP-1300 should assure that it is used in such an environment.

Electromagnetic Emissions: (IEC60601-1-2)

Emission Test

Compliance

Electromagnetic Environment

 

 

 

RF emission

Group 1

The HBP-1300 uses RF energy only for internal functions.

CISPR 11

 

Therefore, this RF emission is extremely weak and there is

 

 

little chance of it creating any kind of interference

 

 

whatsoever with nearby electronic equipment.

 

 

 

RF emissions

Class B

The HBP-1300 is suitable for use in all establishments,

CISPR 11

 

including domestic establishments and those directly

 

 

connected to the public low voltage power supply network

Harmonic emissions

Class A

that supplies buildings used for domestic purposes.

IEC 61000-3-2

 

 

 

 

 

 

Voltage fluctuations/flicker

Complies

 

IEC 61000-3-3

 

 

 

 

 

Electromagnetic Immunity:

 

 

(IEC60601-1-2)

 

 

Immunity test

IEC60601-1-2

Compliance

Electromagnetic

test level

level

environment - guidance

 

 

 

 

 

Electrostatic discharge

±6 kV contact

±6 kV contact

Floors should be wood, concrete or

(ESD)

±8 kV air

±8 kV air

ceramic tile. If floors are covered with

IEC 61000-4-2

 

 

synthetic material, the relative

 

 

 

humidity should be at least 30 %.

 

 

 

 

Electric fast transient/

±2 kV for

±2 kV for

Mains power quality should be that of

burst

power supply lines

power supply lines

a typical commercial or hospital

IEC 61000-4-4

±1 kV for input/

±1 kV for input/

environment.

 

output lines

output lines

 

 

 

 

 

Surge

±1 kV

±1 kV

Mains power quality should be that of

IEC 61000-4-5

differential mode

differential mode

a typical commercial or hospital

 

±2 kV

±2 kV

environment.

 

common mode

common mode

 

 

 

 

 

Voltage dips, short

<5 % UT

<5 % UT

Mains power quality should be that of

interruptions and

for 0.5 cycle

for 0.5 cycle

a typical commercial or hospital

voltage variations on

 

 

environment. If the user of the

40 % UT

40 % UT

power supply input lines

HBP-1300 requires continued

IEC 61000-4-11

for 5 cycles

for 5 cycles

operation during power mains

 

 

 

70 % UT

70 % UT

interruptions, it is recommended that

 

for 25 cycles

for 25 cycles

the HBP-1300 be powered from an

 

uninterruptible power supply or a

 

 

 

 

<5 % UT

<5 % UT

 

battery.

 

for 5 sec.

for 5 sec.

 

 

 

 

 

Power frequency

3 A/m

3 A/m

Power frequency magnetic fields

(50/60 Hz) magnetic

 

 

should be at levels characteristic of a

field

 

 

typical location in a typical commercial

IEC 61000-4-8

 

 

or hospital environment.

 

 

 

 

Note: UT is the a.c. mains voltage prior to application of the test level.