Check before Use

Before conducting safety checks, be sure to implement the items in the “Cleaning of the Device” and “Accessory Care” sections. (page 24)

Before turning on the power

Before turning on the power, check for the following

External appearance

￿The device or accessories are not deformed due to falling or other impact.

￿The device is not dirty.

￿The device is not wet.

AC adapter

￿The AC adapter is firmly connected to the connector on the device.

￿There are no heavy objects lying on the AC adapter cable.

￿The AC adapter cable is not damaged (core-wire exposure, breaks, etc.).

When turning on the power

When turning on the power, check the following

Display/lamp

￿When the [START/STOP] button is pressed to turn on the power, the screen below appears and the alarm lamp lights (page 13).

The alarm lamp is lit up.

EN

After turning on the power

After turning on the power, check for the following

External appearance

￿There is no smoke or odor coming from the device.

￿The device is not making any unusual noises.

Buttons

￿Press each button and check that it works.

Non-invasive blood pressure (NIBP)

￿Make sure that a suitable OMRON GS CUFF is attached (one that fits the circumference of the patient’s arm).

￿The cuff tube is firmly connected.

￿The person checking the cuff should wrap the cuff around arm, perform cuff measurement and check to see that blood pressure is in the vicinity of normal measurements.

￿While measurement is in progress, bend the relevant arm and move body to halt discharge and during this halt check that cuff pressure does not drop.

25

Page 27
Image 27
Omron HBP-1300 Check before Use, Before turning on the power, When turning on the power, After turning on the power