3.4 Using the Memory Function

3.4 Using the Memory Function

The unit automatically stores up to 60 sets of measurement values (blood pressure and pulse rate).

It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes. If there are only two readings in memory for that period, the average will be based on two readings. If there is one reading in memory for that period, the average will be based on one reading.

Notes:

If the memory is full, the monitor will delete the oldest readings.

When viewing the reading taken without setting the date and time, “-:--” is displayed instead of the date and time.

To View the Average Value

Press the Memory button.

Note: If there are no measurements results stored in the memory, the screen to the right is displayed.

To View the Readings Stored in Memory

1.Press the To View the Average Value button, while the average value is displayed.

The Memory number appears for a second before the pulse rate is displayed. The newest

set is numbered “1”. Alternating date/time display

2.Press the To View the Readings Stored in Memory button repeatedly to view the readings stored in memory.

To Delete All the Values Stored in Memory

When the memory symbol ( To Delete All the Values Stored in Memory) appears, first press the 4.Troubleshooting and Maintenance button. Then while holding it down, press the START/STOP button simultaneously for about 2 - 3 seconds.

First Second

Note: You cannot partially delete the stored readings.

4.Troubleshooting and Maintenance

4.1The Icons and Error Messages

Error Display

Cause

Remedy

 

 

 

 

 

Remove the wrist cuff. Wait

 

 

2 - 3 minutes and then take

 

Irregular heartbeats

another measurement.

 

are detected.

Repeat the steps in section

 

 

3.3. If this error continues to

 

 

appear, contact your doctor.

 

 

 

 

Movement during

Carefully read and repeat the

 

measurement.

steps in section 3.3.

 

 

 

 

The batteries are

You should replace them with

Blink

new ones ahead of time.

low.

Refer to section 2.1.

 

 

 

 

 

 

The batteries are

You should replace them with

Lit

new ones at once.

exhausted.

Refer to section 2.1.

 

 

 

 

 

 

Wrist cuff not

Apply the wrist cuff correctly.

 

applied correctly.

Refer to section 3.1.

 

 

 

 

 

Repeat Measurement while

 

Movement during

remaining still and refraining

 

from talking during the

 

measurement.

measurement.

 

 

 

 

Refer to section 3.3.

 

 

 

 

 

Apply the wrist cuff correctly

 

Wrist cuff not

and repeat measurement

 

applied correctly, or

while remaining still and

 

movement during

refraining from talking during

 

measurement.

the measurement.

 

 

Refer to section 3.1 and 3.3.

 

 

 

 

Device error.

Contact your OMRON retail

 

outlet or distributor.

 

 

 

 

 

Note: The irregular heartbeat symbol ( 4.2 Troubleshooting ) may also be displayed with error messages.

4.2 Troubleshooting

Problem

Cause

Remedy

 

 

 

 

The wrist cuff is not

Measure while in the

 

at heart level.

correct posture.

 

Refer to section 3.2.

 

 

 

 

 

 

The cuff is not

Wrap the cuff correctly.

 

wrapped snugly

The reading is

Refer to section 3.1.

around the wrist.

 

extremely low

 

 

The arms and

Relax and try taking the

(or high).

 

shoulders are

measurement again.

 

tense.

Refer to section 3.3.

 

 

 

 

Movement or

Remain still and do not

 

talking during

talk during Measurement.

 

measurement.

Refer to section 3.3.

 

 

 

Wrist cuff pressure

Air is leaking from

Consult your OMRON

does not rise.

the wrist cuff.

retail outlet or distributor.

 

 

 

 

 

Apply the cuff correctly so

Wrist cuff deflates

The wrist cuff is

that it is firmly wrapped

too soon.

loose.

around the wrist. Refer to

 

 

section 3.1.

 

 

 

 

 

Blood pressure readings

 

 

constantly vary with time

The blood pressure is different

of day and how relaxed

each time. The reading is

you are. Take several

extremely low (or high).

deep breaths and try to

 

 

remain relaxed before

 

 

taking a measurement.

 

 

 

The unit loses

 

Replace the batteries with

power during

The batteries are

measurement.

new ones.

exhausted.

 

Refer to section 2.1.

 

 

Nothing happens

 

 

when you press

The batteries have

Insert the batteries with

the buttons.

been inserted

the correct (+/-) polarity.

 

incorrectly.

Refer to section 2.1.

 

 

 

 

• Press the START/STOP button and repeat

 

measurement.

 

Other problems.

• If the problem continues, try replacing the

batteries with new ones.

 

 

If this still does not solve the problem, contact

 

your OMRON retail outlet or distributor.

 

 

 

4.3 Maintenance

To protect your unit from damage, please observe the following:

•Do not subject the main unit and the cuff to extreme temperatures, humidity, moisture or direct sunlight.

•Do not disassemble the unit.

•Do not subject the unit to strong shocks or vibrations (for example, dropping the unit on the floor).

•Do not use volatile liquids to clean the main unit. •Do not wash the cuff or immerse it in water. •Do not use petrol, thinners or similar solvents to clean the cuff.

•Do not carry out repairs of any kind yourself. If a defect occurs, consult your OMRON retail outlet or distributor as mentioned on the packaging.

•The unit should be cleaned with a soft, dry cloth. •Use a soft, moistened cloth and soap to clean the cuff.

•Keep the unit in its storage case when not in use.

•Fold the cuff into the storage case.

Do not store the unit in the following situations:

•If the unit is wet.

•Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapours.

•Locations exposed to vibrations, shocks or where it will be at an angle.

Calibration and Service

•The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.

•It is generally recommended to have the unit inspected every two years to ensure correct functioning and accuracy. Please consult your authorised OMRON dealer or the OMRON Customer Service at the address given on the packaging or attached literature.

5. Technical Data

Product Description

Wrist Blood Pressure Monitor

Model

OMRON RS3 (HEM-6130-E)

Display

LCD Digital Display

Measurement Method

Oscillometric method

Measurement Range

Pressure: 0 to 299 mmHg

 

Pulse: 40 to 180/min.

Accuracy

Pressure: ±3 mmHg

 

Pulse: ±5% of display reading

Inflation

Automatic inflation by pump

Deflation

Automatic rapid deflation

Memory

60 Measurements

Power Source

2 “AAA” alkaline (LR03) batteries 1.5V

Battery Life

Approx. 300 measurements with new

 

alkaline batteries at a room temperature of

 

23°C

Applied Part

= Type B

Protection Against

Internally powered ME equipment

Electric Shock

 

Operating temperature/

+10 to +40°C / Maximum: 30 to 85% RH

Humidity

 

Storage temperature/

-20 to +60°C / Maximum: 10 to 95% RH /

Humidity

700 to 1060hPa

Console Weight

Approximately 101g without batteries

Outer Dimensions

Approximately 78 (w) mm x 60 (h) mm x

 

21 (d) mm (without the wrist cuff)

Measurable

Approximately 13.5 to 21.5 cm

circumference

 

Cuff Material

Nylon and polyester

Package Content

Main unit, storage case, battery set,

 

instruction manual, guarantee card

Note: Subject to technical modification without prior notice.

This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive).

This blood pressure monitor is designed according to the European Standard EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.

This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the Pressure Sensor, is produced in Japan.

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.

This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions.

Nevertheless, special precautions need to be observed:

Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual.

Documentation is also available at www.omron-healthcare.com.

Correct Disposal of This Product

(Waste Electrical & Electronic Equipment)

This marking shown on the product or its

literature, indicates that it should not be disposed of, with other household wastes at the

end of its working life.

To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.

Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.

Manufacturer

OMRON HEALTHCARE Co., Ltd.

 

53, Kunotsubo, Terado-cho, Muko, Kyoto,

 

617-0002 JAPAN

 

 

EU-representative

OMRON HEALTHCARE EUROPE B.V.

 

Scorpius 33, 2132 LR Hoofddorp

 

THE NETHERLANDS

 

www.omron-healthcare.com

 

 

Production facility

OMRON (DALIAN) CO., LTD.

 

Dalian, CHINA

Subsidiary

OMRON HEALTHCARE UK LTD.

 

Opal Drive, Fox Milne, Milton Keynes, MK15 0DG

 

U.K.

 

 

 

OMRON MEDIZINTECHNIK

 

HANDELSGESELLSCHAFT mbH

 

John-Deere-Str. 81a, 68163 Mannheim, GERMANY

 

www.omron-medizintechnik.de

 

 

 

OMRON SANTÉ FRANCE SAS

 

14, rue de Lisbonne, 93561

 

Rosny-sous-Bois Cedex, FRANCE

 

 

 

Made in China

Page 2
Image 2
Omron R53 Using the Memory Function, To View the Average Value, To View the Readings Stored in Memory, Troubleshooting