Omron Healthcare 705IT Hints on taking blood pressure readings, Failure, causes and rectification

Models: 705IT

1 12
Download 12 pages 44.31 Kb
Page 9
Image 9

Hints on taking blood pressure readings

16

Before measuring, stretch your back and sit up straight, breath normally 5-6 times, relax your shoulders, arms and entire body.

Slightly bend and support your elbow (e.g. on a table).

Do not move, talk or touch the device during measurement.

Stress influences your blood pressure, so try to relax before taking a measurement.

To compare results, measurements must always be taken from the same arm.

Do not place the cuff over thick clothes and do not roll up your sleeve if it is too tight.

Align the cuff with the height of your heart (breast height). If necessary use a pillow or cushion.

Coldness raises your blood pressure. It is recommended to measure at room temperature (+20°C).

Relax before measurement. If necessary go to the bathroom before measurement.

Do not measure directly after bathing.

If you wish to interrupt measurement prematurely, press the (O/I) button. Inflation is stopped and the air is automatically released from the cuff.

Blood pressure measurements taken in a moving vehicle may be influenced by vibration.

If an error has occurred during measurement E is displayed. Please refer to the next chapter.

The monitor should be switched off after measurement to avoid running down the batteries unnecessarily. If you forget to do so, the monitor switches itself off automatically after 5 minutes.

English

Failure, causes and rectification

Failure and their possible causes

Rectification

Incorrect measurement is indicated by E

 

 

 

 

1.

A correct reading could not be obtained

1.

Repeat the measurement keeping

 

because measurement was disturbed by

 

perfectly still. Do not move your arm and

 

movement of the body.

 

do not speak.

2.

The cuff is not fitted correctly.

2.

Check that the cuff is correctly fitted.

 

 

 

 

 

 

 

 

Then repeat the measurement.

3.

Your clothing has constricted your

3.

Remove the item of clothing which

 

blood flow.

 

 

 

caused the constriction.

4.

There is still air in the cuff when the

4.

The unit may be defective. Please have

 

monitor is switched off.

 

it examined and, if necessary, repaired

 

 

 

 

 

 

 

 

by OMRON.

 

 

 

 

The (START

 

) button was pressed

Wait for the ( ) symbol indicating

 

 

before the (

 

) symbol was displayed.

readiness before you press the (START

 

)

 

 

 

 

 

 

 

 

 

 

 

button.

 

 

The blood pressure values displayed are

Refer to the instructions and then repeat

extremely low or high, or they are

the measurement.

implausible.

 

 

 

 

 

 

 

 

 

 

 

The display does not light up when the

 

 

 

 

(START

 

) button is pressed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.

The batteries are exhausted.

1.

Check the batteries and, if necessary, fit

 

 

 

 

 

 

 

 

4 new batteries.

2.

The +/- poles of the batteries have been

2.

Reinsert the batteries with the +/- poles

 

reversed.

 

 

 

the right way round.

3.

The battery contacts are dirty.

3.

Clean the battery contacts with a dry

 

 

 

 

 

 

 

 

cloth.

 

 

The batteries are weak or exhausted.

Fit new batteries.

 

 

The cuff pressure does not rise

Check that the air tube is properly

although the pump motor can be heard.

connected to the monitor. Push the tube

 

 

 

 

 

 

 

connector firmly into the socket.

 

problem with mem. function

Contact your Omron retailer or distributor.

Er

 

 

 

 

 

 

 

 

17

Page 9
Image 9
Omron Healthcare 705IT instruction manual Hints on taking blood pressure readings, Failure, causes and rectification