Omron Healthcare HEM-711 HOW to USE the Memory Function, Instructions for special conditions

Models: HEM-711

1 10
Download 10 pages 62.72 Kb
Page 7
Image 7

HOW TO TAKE A READING

Instructions for special conditions:

If your systolic pressure is known to be more than 220 mmHg, push and hold the START button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure.

NOTE: Do not apply more pressure than necessary.

The monitor will not inflate above 300 mmHg.

• If you want to stop the

 

measurement, push the ON/OFF

 

button. The monitor will stop

START

inflating and start deflating

rapidly, then the monitor will

 

power off.

 

 

Continue to press button

HOW TO USE THE MEMORY FUNCTION

This monitor has a memory capable of storing fourteen sets of readings. Every time you complete the measurement, the monitor automatically stores blood pressure and pulse rate.

Push the button to recall stored readings while the Heart Symbol ( ) is displayed.

When the memory is holding fourteen sets of readings, the oldest set will be deleted to store a new set.

Last reading

Value before

Value of fourteen

the last reading

readings before

To delete stored readings, press the button and the Start button at once, then all the readings will be deleted.

NOTE: You can not delete the stored readings partially.

START

Press these two buttons

Display of memory deletion

at the same time.

 

12

13

Page 7
Image 7
Omron Healthcare HEM-711 instruction manual HOW to USE the Memory Function, Instructions for special conditions