HOW TO USE THE OPTIONAL AC ADAPTER

(ACCESSORY ITEM SOLD SEPARATELY)

To avoid potential damage to the monitor, use only the exclusive AC adapter (Model HEM-ADPT1) which can be purchased from:

Omron Healthcare, Inc.

300 Lakeview Parkway

Vernon Hills, IL 60061

Toll Free Phone #: 1-800-634-4350

To connect AC adapter

1. Insert the AC adapter plug Plug into the jack on the side of the

monitor.

2. Insert the AC adapter plug

into a 120V AC outlet (50 - 60

cycles).Jack

Option AC Adapter

To remove the AC adapter

To remove the AC adapter, disconnect the adapter plug from the AC outlet first and then disconnect the cord from the monitor’s jack.

NOTE: The monitor is designed not to draw power from the batteries when the AC adapter is in use.

TROUBLESHOOTING

Error Indicator

Cause

Correction

 

Cuff under-inflated.

Turn monitor off, wait 2-3 minutes

 

and take another measurement.

 

 

 

 

Repeat steps listed under "How to

 

 

Take a Reading." If "EE" is displayed

 

 

again, take another measurement

 

 

 

 

by pressing and holding the START

 

Movement during

button as instruction page 12 under

 

"How to Take a Reading", instructions

 

measurement.

for special conditions.

Cuff over-inflated (more than

Turn the monitor off, carefully read

and repeat steps listed under "How

300 mmHg).

to Take a Reading."

 

Battery voltage is excessively

Replace all four "AA" batteries with

low.

new batteries.

Problem caused by abnormal

Contact OMRON Customer Service

memory function.

Toll Free 1-800-634-4350

If you continue to have difficulty obtaining a reading, call

OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350.

Please tell the representative you have an OMRON Model HEM-711.