18

Light Source

Irradiance Level

Standard Panel-Light:

30µW/cm2/nm

Neonatal Panel-Light:

55µW/cm2/nm

Intensity Ratio

Standard Panel and Neonatal Panel : > .4 (minimum to maximum)

Audible Noise

< 60 dB(A). Measured in accordance with IEC 60601-2-50.

Standards Compliance

This device is designed to conform to the following standards:

IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment

IEC 60601-2-50 Requirements for the Safety of Infant Phototherapy Equipment

Electromagnetic Compatibility: EN 60601-1-2, 2nd edition.

Electrical Requirements

AC Power

100-240 VAC, 50/60 Hz, 1.0 A

Type of Protection Against Electrical Shock

Class I Equipment

 

 

Degree of Protection Against Electrical Shock

Type BF Applied Part

 

 

Degree of Protection Against Ingress of Water

Ordinary Protection, IPX0

Mode of Operation

Continuous

Disposal

Dispose of this device in accordance with local regulations.

BiliTx Parents’ Manual

Page 20