Philips M2926A Specifications, Monitor M2922A, Transducers, Remote Event Marker 15249A, Parameter

Models: M2922A M2925A

1 8
Download 8 pages 22.87 Kb
Page 4
Image 4
Specifications

Specifications

Monitor (M2922A)

Federal Law restricts this device to use by or on the order of a physician.1

Standards Compliance

The FM-2 Portable Antepartum Fetal Monitor complies with the requirements of the Medical Device Directive 93/42/EEC and the following safety standards:

EN60601-1/IEC60601-1, EN60601-1-1, EN60101-2-37

IEC 1157

IEC 60601-1

UL2601-1

CAN/CSA C22.2#601.1-M90

Class II equipment, double insulated

Continuous operation

Parameter

 

 

Input

 

Resulting

 

Connector

 

Isolation

 

 

 

 

 

 

 

 

 

US

 

 

BF

 

BF

 

 

 

 

 

 

Toco

 

 

BF

 

BF

 

 

 

 

Electromagnetic Compatibility

 

Complies with EN60601-1-2

 

Environmental

 

 

 

 

Operating Temp:

 

-10oC to 45oC

 

 

 

(50oF to 110oF)

Storage Temp:

 

-20oC to 60oC

 

 

 

(-29oF to 175oF)

Relative Humidity:

20% to 90% non

 

 

 

condensing

 

Altitude:

 

0-3048m (10,000 ft)

Dimensions

 

 

 

 

Height:

 

15.24 cm (6.0 inches)

Width:

 

26.4 cm (10.4 inches)

Depth:

 

13.7 cm (5.4 inches)

Weight:

 

2.27 kg (5lb)

 

Power - Battery

 

 

 

 

Type:

 

Lead acid, rechargeable.

Operation:

 

minimum 2-hours fully

 

 

 

charged at 10oC, longer

 

 

 

at higher temperature.

Recharge during

 

Maximum 14 hours full

monitoring:

 

recharge during monitor-

 

 

 

ing (typical 11 hours).

Recharge without

Maximum 8 hours to

monitoring:

full recharge when no

 

monitoring occurs (typical

 

4.75 hours).

Low power warning: Approximately 30 minutes

 

at 25oC.

Power - External Power Supply

AC-powered

110-120V~, 50-60Hz,

 

0.15A

 

or

 

220-240V~, 50-60Hz,

 

0.08A

Display

 

Backlight:

Cold cathode fluorescent

Active area:

3.78” x 2.83”

 

(320 x 240 dots)

Resolution:

85 dots per inch

Contrast:

10:1

Ultrasound

 

Isata:

< 4mW

BPM range:

 

Paper scale:

USA paper: 30-240 BPM,

 

International Paper: 50-210

 

BPM.

Ultrasound

 

processing range:

50-240 BPM.

Accuracy:

± 1% ± 1 BPM.

Leakage current:

10 A @ 264 VAC applied

 

to transducer.

Isolation:

>4 kV RMS, Type BF

 

applied part.

Toco

 

Range:

0-100 relative units

Resolution:

1 count

Accuracy:

± 1% ± 1 relative unit

Leakage:

10 A @ 264 VAC applied

 

to transducer

Isolation:

>4 kV RMS, Type BF

 

applied part

Transducers

Doppler Ultrasound FHR (M1356A)

Weight:

185 g (6.5 oz)

Cable length:

2.5m (8ft 2in)

Diameter:

75mm (2.95in)

Latex free

 

External Toco (M1355A)

Transducer Weight: 180 g (6.3 oz)

Cable length: 2.5m (8ft 2in)

Diameter: 75mm (2.95in)

Latex free

Remote Event Marker (15249A)

1.USA only

Length: 2.8m (9ft 2in)

4

Page 4
Image 4
Philips M2926A manual Specifications, Monitor M2922A, Transducers, Remote Event Marker 15249A, Parameter, Input, Resulting