battery is available, remove the installed battery for five seconds, then reinsert it

and run a battery insertion self-test.

If the problem continues, do not use the OnSite. Attend to the patient, providing

CPR if needed, until Emergency Medical Services Personnel arrive.

TROUBLESHOOTING WHILE THE ONSITE IS IN USE

(green Ready light is solid)

OnSite tells you:

... to replace the battery immediately

... there is no cartridge installed, and

... to insert a pads cartridge

... to press the pads firmly to the skin

... to make sure the pads have been removed from the liner

... the pads should not be touching the patient’s clothing.

... to insert new pads cartridge

possible cause

The battery is nearly depleted. The OnSite will turn off if a new battery is not inserted.

The pads cartridge has been removed.

The pads cartridge has been damaged.

The pads are not properly applied to the patient.

The pads are not making good contact with the patient's bare chest because of moisture or excessive hair.

The pads are touching each other.

The pads may not have been removed from the liner or may be on the patient’s clothing.

The pads cartridge has been opened and the pads peeled off the liner, but the pads have not been successfully attached to the patient. There may be a problem with the pads cartridge.

recommended action

Replace the battery with a new battery immediately.

Insert a new pads cartridge.

Make sure that the pads are sticking completely to the patient’s skin.

If the pads are not sticking, dry the patient's chest and shave or clip any excessive chest hair.

Reposition the pads.

Make sure the pads are not on the liner or the patient’s clothing.

If the voice instruction continues after you do these things, insert another pads cartridge.

Replace the damaged pads cartridge. Pull up the handle on the cartridge cover, and replace pads on patient with new pads to continue with the rescue.

PHILIPS MEDICAL SYSTEMS

G-2

Page 58
Image 58
Philips M5066A owner manual Troubleshooting While the Onsite is in USE, Green Ready light is solid OnSite tells you

M5066A specifications

The Philips M5066A is a state-of-the-art multi-parameter patient monitor designed to enhance healthcare delivery in various clinical settings. This versatile monitoring solution is widely utilized in hospitals, especially in critical care units and during transport, owing to its robust features and reliable performance.

One of the main features of the M5066A is its ability to monitor multiple vital signs simultaneously, including ECG, heart rate, invasive and non-invasive blood pressure, respiration rate, temperature, and saturation levels. This multi-parameter capability allows healthcare professionals to gain a comprehensive view of a patient’s health status, enabling timely interventions and improved patient outcomes.

The M5066A incorporates advanced technologies such as Philips’ proprietary IntelliVue technology, which enhances the accuracy and reliability of vital sign measurements. The monitor uses sophisticated algorithms to filter out noise and interferences, ensuring that the data presented to caregivers is both precise and actionable. The device also supports a range of connectivity options, allowing for seamless integration with hospital information systems and electronic medical records (EMRs). This connectivity facilitates the easy transfer and management of patient data, promoting better coordination of care.

Another significant characteristic of the M5066A is its user-friendly interface. The large, high-resolution display provides clear visuals of vital sign trends and parameters, enabling quick assessments by healthcare professionals. Customizable alarms and alerts can be programmed to suit specific patient needs, ensuring that caregivers are notified of critical changes in a timely manner.

The M5066A is also designed with patient comfort and safety in mind. The device's compact and lightweight design facilitates easy transport, while its durable construction ensures reliability in demanding environments. Furthermore, the monitor's battery backup allows for continuous monitoring even during power outages or when moving patients between departments.

In conclusion, the Philips M5066A multi-parameter patient monitor is a sophisticated tool that enhances patient monitoring by providing accurate, real-time data. Its advanced technologies, comprehensive monitoring capabilities, user-friendly interface, and focus on safety and comfort make it an invaluable asset in any clinical setting, helping healthcare providers deliver the highest standard of care.