PHILIPS MEDICAL SYSTEMS
E-1
E
E TECHNICAL INFORMATION

HEARTSTART ONSITE DEFIBRILLATOR SPECIFICATIONS

The specifications provided in the following tables are nominal values. Additional
information can be found in the Technical Reference Manuals for HeartStart
Automated External Defibrillators, located online at www.philips.com/
productdocumentation.

PHYSICAL

ENVIRONMENTAL

category specifications
size 2.80” H x 7.40” D x 8.30” W (7.1cm H x 19cm D x 21cm W).
weight Approximately 3.3 lbs (1.5 kg) with battery and pads cartridge installed.
category specifications
temperature and
relative humidity
Operating (battery and pads cartridge installed):
32° to 122° F (0° to 50° C)
0% to 95% RH (non-condensing).
Standby (battery and pads cartridge installed):
50° to 109° F (10° to 43° C)
10% to 75% RH (non-condensing).
Storage/shipping (with battery and pads cartridge):
-4° to 140° F (-20° to 60° C) for up to 2 days
0% to 85% RH (non-condensing)
altitude Operates at 0 to 15,000 feet; can be stored at up to 8,500 feet, in standby
mode.
shock/drop abuse
tolerance
Withstands 1 meter drop to any edge, corner, or surface.
vibration Operating: meets EN1789 random, road ambulance.
Standby: meets EN1789 swept sine, road ambulance.