11
Calibrating
Locking Calibration SettingsThe generator uses an internal hardware jumper to enable and disable changes to calibration
settings. If the Options > Calibration option is available, the jumper is set to enable re-calibration.
To prevent users from changing calibration settings:
1. Turn the generator off, and remove the back cover.
2. Locate the J10 jumper, which is near the lower-right corner of the board.
3. Connect the jumper to only one pin to disable user calibration. Connect both bins to enable
user calibration.
4. Install the back cover.
Adjusting Horizontal Scanning FrequencySelect Test > Freq Shift, and then press the +/- buttons to increase or decrease the horizontal
scanning frequency of the current video format +/- 10% in 2% increments. The ^ character indicates
the percent deviation. The bottom number is the actual frequency.
_____Freq. Shift
-10%_-----+-----_+10
__________^
__H= +31.469E+03