TROUBLESHOOTINGTITLE

 

Display

 

 

Meaning

Action

 

 

 

 

 

Lens cover is not

Attach the lens cover,

 

 

 

 

 

 

 

 

 

 

on Probe.

then measure.

 

 

 

 

 

 

 

 

 

 

 

 

Temperature of the

Make sure unit

 

 

 

 

 

 

 

 

 

unit is unstable.

 

 

 

 

 

 

remains at room

 

 

 

 

 

 

 

 

 

 

 

The temperature of the

temperature where

 

 

 

 

 

measurement will

 

 

 

 

 

unit is 50°F or below

 

 

 

 

 

take place. Allow 30

 

 

 

 

 

 

 

 

 

which is outside

 

 

 

 

 

minutes to stabilize

 

 

 

 

 

the operating range.

 

 

 

 

 

before using.

 

 

 

 

 

 

 

 

 

 

 

 

Normal range:

 

 

 

 

 

The temperature of the

 

 

 

 

 

unit is 104°F or above

(50°F to 104°F)

 

 

 

 

 

 

 

 

 

which is outside the

(10°C to 40°C)

 

 

 

 

 

operating range.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Turn on the unit

 

 

 

 

 

 

again. If "Er.1"

 

 

 

 

 

 

is displayed again,

 

 

 

 

 

 

the unit may

 

 

 

 

 

May require service.

require service.

 

 

 

 

 

Call Customer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Service Dept.

 

 

 

 

 

 

 

 

 

 

 

 

1-800-634-4350

 

 

 

 

 

 

 

 

 

 

 

 

The measured

Probe may not

 

 

 

 

 

temperature was

have been properly

 

 

 

 

 

positioned in ear.

 

 

 

 

 

above the range

 

 

 

 

 

The probe must

 

 

 

 

 

of the thermometer.

 

 

 

 

 

be snug against

 

 

 

 

 

 

 

 

 

 

 

The measured

opening of ear.

 

 

 

 

 

Be sure lens cover

 

 

 

 

 

temperature was

 

 

 

 

 

below the range

is clean and ear

 

 

 

 

 

of the thermometer.

canal is not blocked

 

 

 

 

 

with earwax.

 

 

 

 

 

 