Problem

Possible Cause

Display error

You have experienced

message: High

excessively high leak levels

leak in last

during the night.

session

 

Solution

Check that your air tubing is connected properly and that your mask does not leak excessively.

Use the mask-fitting feature to help you to fit your mask properly.

If this message appears again, contact your clinician.

Display error

You have experienced

message: HIGH

excessively high leak levels for

LEAK!!!

more than 20 seconds.

Adjust Mask

 

Check that your air tubing is connected properly and that your mask does not leak excessively.

Use the mask-fitting feature to help you to fit your mask properly.

If this message appears again, contact your clinician.

Display error

Component failure.

Return your S8 Elite II for

message:

 

servicing.

Exxxx (where xxxx

 

 

defines an error)

 

 

 

 

 

The following message is displayed on the LCD after you try to update settings using the Data Card:

Card Error

Remove Card

The Data Card is not inserted correctly.

The Data Card is only partly inserted.

Ensure that the Data Card is inserted with the arrow facing up and pointing towards the device. Ensure that the device is switched on and in standby mode before you insert the Data Card.

Ensure that the Data Card is inserted into the slot on the Data Card module as far as it will go.

You may have removed the Data Card before settings were copied to the S8 Elite II.

The card is not a ResScan Data Card.

Reinsert the Data Card and wait for the

Settings Success

Remove Card

message to appear on the LCD.

Only use a ResScan Data Card.

Troubleshooting 29

Page 35
Image 35
ResMed 338425/1 manual Leak