Noise Level

AutoSet Spirit II—CPAP mode, 10 cm H2O set pressure*

Measured total A-weighted sound pressure level

24 dBA

(dB re 20 μPa)

 

 

 

Uncertainty (dB re 20 μPa)

2 dBA

Measured values are determined according to the noise test procedure given in ISO 17510-1.

Note: The sum of a measured noise emission value and its associated uncertainty represents an upper boundary of the range of values that is likely to occur in measurements.

*Declared dual number noise emission values in accordance with ISO 4871.

Accessories

Data Card

Dimensions (L x W x D): 65.04 mm x 22.15 mm x 0.85 mm

Data Card Module

Dimensions (H x W x D): 49.7 mm x 42.2 mm x 12 mm

All Devices (where applicable)

Housing Construction: Flame retardant engineering thermoplastic

Environmental Conditions

Operating Temperature: +5°C to +36°C

Operating Humidity: 10%–95% non-condensing

Storage and Transport Temperature: -20°C to +60°C

Storage and Transport Humidity: 10%–95% non-condensing

Operating Altitude: sea level to 2591 m

Electromagnetic Compatibility

Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial and light industry environments. For further details, see “Guidance and Manufacturer’s Declaration — Electromagnetic Emissions and Immunity” on page 34.

Note: The manufacturer reserves the right to change these specifications without notice.

System Specifications

32

Page 38
Image 38
ResMed 338425/1 manual All Devices where applicable, Data Card Module