Solution
Ensure the power cable is connected and the power outlet (if available) is on.
Fully insert the DC plug.

Weekly:

1.Remove the air tubing from the S9 device and the mask.

2.Wash the air tubing in warm water using mild detergent.

3.Rinse thoroughly, hang, and allow to dry.

4.Before next use, reconnect the air tubing to the air outlet and mask. Monthly:

1.Wipe the exterior of the S9 with a damp cloth and mild detergent.

2.Check the air filter for holes and blockage by dirt or dust. Replace the air filter if necessary. Replacing the air filter:

Replace the air filter every six months (or more often if necessary).

1.Remove the air filter cover from the back of the S9 device.

2.Remove and discard the old air filter.

3.Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover.

4.Replace the air filter cover.

Notes:

Ensure the air filter and air filter cover are fitted at all times.

Do not wash the air filter. The air filter is not washable or reusable.

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.

Problem/Possible cause

No display

Power is not connected.

The DC plug is partially inserted into the back of the device.

Insufficient air delivered from the device

Ramp time is in use.

Wait for air pressure to build up or change ramp time.

Air filter is dirty.

Replace air filter.

Air tubing is not connected properly.

Check air tubing.

Air tubing is blocked, pinched or

Unblock or free the air tubing. Check the air tubing for

punctured.

punctures.