ResMed H4i Technical Specifications, Humidifier Performance, Symbols which appear on the device

Models: H4i

1 14
Download 14 pages 58.77 Kb
Page 11
Image 11
Technical Specifications

Technical Specifications

Maximum heater plate temperature

185°F (85°C)

Temperature cut-out

199°F (93°C)

Maximum gas temperature

106°F (41°C)

Dimensions (H x W x D)

Docking station and water chamber: 5.0” x 6.7” x 8.3” (130 mm x

 

170 mm x 212 mm)

Weight

Docking station and unfilled water chamber: 1.6 lb (732g)

Water capacity

To maximum fill line: 390 mL

Power supply

Input range for ResMed S8 flow generator with H4i: 100–240V,

 

50/60Hz; 110V, 400Hz; 2.5A <140 VA (110 W) (max. power

 

consumption) Instantaneous peak power consumption <340 VA*

 

Maximum heater element power: 85 W

Docking station

Flame retardant engineering thermoplastic, plated cast aluminium

Water chamber

Injection molded plastic, stainless steel and silicon seal

Operating temperature

+41°F to +95°F (+5°C to +35°C)

Operating humidity

10–95% non-condensing

Storage and transport temperature

-4°F to +140°F (-20°C to +60°C)

Storage and transport humidity

10–95% non-condensing

IEC 60601-1 classification

Class II (double insulation), Type CF

Electromagnetic compatibility

Please refer to your flow generator user's manual for details

*Actual average power consumption will vary depending on factors such as temperature setting, set pressure, the use of accessories, height above sea level and ambient temperature.

Humidifier Performance

The following settings have been tested at 22°C, 40% RH ambient conditions:

CPAP Mask Pressure

 

RH Output %

Nominal System Output AH*, BTPS

cm H2O

Setting 3

Setting 6

Setting 3

Setting 6

4

>95

>95

16

20

 

 

 

 

 

10

>85

>95

14

18

 

 

 

 

 

20

>75

>85

13

16

 

 

 

 

 

*AH- Absolute Humidity in mg/L.

BTPS- Body Temperature Pressure Saturated.

Note: The manufacturer reserves the right to change these specifications without notice.

Symbols which appear on the device

Attention, consult accompanying documents;

Type CF equipment;

Class II equipment;

Hot surface;

Do not touch;

Humidifier Performance Maximum water level.