Operating pressure

4 to 20 cm H2O

range

 

Dimensions

4.6” x 6.5” x 5.7”

(H x W x D)

(112 mm x 164 mm x 145 mm)

Weight

2.9lb (1.3 kg)

Power Supply

Input range for S8 Escape II with HumidAire 3i:100–240V, 50/60Hz;

 

110V, 400Hz; 2.5A < 140VA (110W) (maximum power consumption).

 

Instantaneous peak power consumption <340VA.

 

Refer to the DC-12 converter instructions for DC ratings. Only use

 

the ResMed DC-12 converter for DC input.

 

Typical power consumption at 20 cm H20 is 34VA or 20W.

Air Filter

Two-layered, powder-bonded, polyester non-woven fiber

Air Tubing

Flexible plastic, 6’6” (2 m)

Air Outlet

The 22 mm conical outlet complies with EN 1281-1

IEC 60601-1

Class II (double insulation), Type CF

Classifications

 

Noise Level

Radiated sound pressure is measured at <30dBA at 10 cm H2O and

 

3’3.36” (1 m) according to ISO17510-1.

Supplemental Oxygen

Recommended maximum supplemental oxygen flow: 4 L/min

Housing Construction

Flame retardant engineering thermoplastic

Operating

+41ºF to +95ºF (+5ºC to +35ºC)

Temperature

 

Operating Humidity

10%–95% non-condensing

Storage and Transport

-4ºF to +140ºF (-20ºC to +60ºC)

Temperature

 

Storage and Transport

10%–95% non-condensing

Humidity

 

Operating Altitude

Sea level to 8500’ (2591 m)

Electromagnetic

Product complies with all applicable electromagnetic compatibility

Compatibility

requirements (EMC) according to IEC60601-1-2, for residential,

 

commercial, and light industry environments.

 

The electromagnetic compatibility tables for these ResMed devices can

 

be found on www.resmed.com, on the Products page under Service

 

& Support. Click on the PDF file for your device.

Note: The manufacturer reserves the right to change these specifications without notice.

Symbols which appear on the device

Attention, consult accompanying documents; Drip proof; Type CF equipment;

Dangerous voltage; Class II equipment; Start/Stop.