Guidance and manufacturer’s declaration—electromagnetic immunity

The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.

Immunity test

IEC60601-1-2

Compliance

Electromagnetic environment—guidance

test level

level

 

 

 

 

 

 

 

 

 

Portable and mobile RF communications

 

 

 

equipment should be used no closer to any part

 

 

 

of the S8 device, including cables, than the

 

 

 

recommended separation distance calculated

 

 

 

from the equation applicable to the frequency of

 

 

 

the transmitter.

 

 

 

Recommended separation distance

Conducted RF

3 Vrms

3 Vrms

d = 1.17 P

IEC 61000-4-6

150 kHz to 80 MHz

 

 

Radiated RF

10 V/m

10 V/m

d = 0.35 P 80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5 GHz

 

 

 

 

 

d = 0.70 P 800 MHz to 2.5 GHz

 

 

 

where P is the maximum output power rating of

 

 

 

the transmitter in watts (W) according to the

 

 

 

transmitter manufacturer and d is the

 

 

 

recommended separation distance in meters (m).

 

 

 

Field strengths from fixed RF transmitters, as

 

 

 

determined by an electromagnetic site survey,a

 

 

 

should be less than the compliance level in each

 

 

 

frequency range.b

 

 

 

Interference may occur in the vicinity of

 

 

 

equipment marked with the following symbol:

 

 

 

 

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the S8 device is used exceeds the applicable RF compliance level above, the S8 device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the S8 device.

bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

ENGLISH

SYSTEM SPECIFICATIONS

33

Page 34
Image 34
ResMed S8 ESCAPE SYSTEM, s8 user manual Recommended separation distance

S8 ESCAPE SYSTEM, s8 specifications

The ResMed S8 Escape is a highly regarded CPAP (Continuous Positive Airway Pressure) device designed for the treatment of obstructive sleep apnea. Known for its reliability and user-friendly features, the S8 Escape is part of ResMed’s S8 series, which combines advanced technologies with an emphasis on patient comfort.

One of the standout features of the S8 Escape is its AutoRamp technology, which helps patients gradually transition into their prescribed pressure settings. Instead of starting at the full pressure immediately, AutoRamp begins at a lower pressure, allowing users to fall asleep comfortably before automatically increasing to the prescribed level. This feature significantly improves the overall user experience as it reduces anxiety associated with the use of CPAP devices.

The S8 Escape also boasts a built-in air filter that enhances air quality and protects the device from dust and other particulates. Additionally, the device is designed for whisper-quiet operation, ensuring that users can sleep soundly without the disruption of noise, which can be a common issue with less advanced models. This creates a more peaceful sleeping environment, both for the user and their bed partner.

Another important characteristic of the S8 Escape is its compact design and lightweight build. It is easily portable, making it an ideal choice for both home use and travel. The device also features a user-friendly interface, with a simple LCD screen that displays essential information such as pressure settings and therapy time. This allows users to track their usage and monitor their therapy progress conveniently.

The S8 Escape offers various pressure settings and is compatible with a wide range of masks, ensuring that users can find the right fit for their individual needs. The device’s advanced monitoring capabilities also help in providing feedback on therapy effectiveness, allowing healthcare professionals to adjust settings if necessary.

In summary, the ResMed S8 Escape is a highly functional and patient-friendly CPAP device, featuring advanced technologies designed to enhance the user experience. Its AutoRamp technology, quiet operation, and compact design make it a popular choice among individuals dealing with obstructive sleep apnea.