Problem/Possible Cause

Solution

 

 

Excessive leak.

Adjust position of mask and headgear.

 

Replace caps if missing from ports on mask.

 

Air tubing may not be connected properly. Connect

 

firmly at both ends.

 

Air tubing may be kinked or punctured. Straighten or

 

replace.

Displays error message: SYSTEM ERRxxxx Call Service! or Exxxx (where xxxx defines an error)

Component failure.

Record error number and call service.

Displays error message: Check tube!! Key if done

Air tubing is loose or blocked.

Check that the air tubing is connected securely to your mask and the air outlet to the front of the S8 Series II. When you have checked the air tubing, press the Start/ Stop key to restart the device. If this does not clear the message, disconnect the power cord and then reconnect it to restart the device.

The following message is displayed on the LCD after you try to update settings or copy data to the Data Card: Card Error Remove Card

Data Card is not inserted correctly.

Check that the Data Card is inserted into the Data Card

 

module with the arrow facing up, while the S8 Series II

 

is switched on and not delivering therapy (standby

 

mode).

The Data Card is only partly inserted.

You may have removed the Data Card before settings were copied to the S8 II.

The card is not a ResScan Data Card.

Ensure that the Data Card is inserted into the Data Card slot as far as it will go.

Reinsert the Data Card and wait for the Settings Success Remove Card message to appear on the LCD.

Only use a ResScan Data Card.

The following message is displayed on the LCD after you try to update settings or copy data to the Data Card: Settings Error Remove Card

There is a data error on the Data

Contact your clinician immediately.

Card.

 

The following message is NOT displayed on the LCD after you try to update the settings using the Data Card: Settings Success Remove Card

The settings were not updated.

Contact your clinician immediately.

Troubleshooting 13

Page 15
Image 15
ResMed S8 AutoScore II, S8 Series II manual Displays error message Check tube!! Key if done

S8 AutoScore II, S8 Series II specifications

The ResMed S8 Series II is a prominent line of CPAP (Continuous Positive Airway Pressure) devices designed to treat obstructive sleep apnea effectively. One of the standout models within this series is the S8 AutoScore II, which has been engineered to adapt to varying patient needs with precision and comfort.

A key feature of the S8 AutoScore II is its AutoSet technology, which continuously monitors a patient's breathing patterns throughout the night. This innovation allows the device to automatically adjust the pressure delivered based on real-time data. As a result, users benefit from personalized therapy, ensuring that they receive the optimal level of air pressure needed at any given moment while they sleep.

The S8 AutoScore II also boasts an intelligent ResMed algorithm, which enhances its ability to differentiate between obstructive and central apneas. This specificity not only improves treatment efficacy but also contributes to increased patient comfort, minimizing the chances of unnecessary pressure increases that could lead to discomfort or disturbances during sleep.

Another notable aspect of the S8 AutoScore II is its user-friendly interface. The device features an intuitive LCD screen that displays essential information and settings, making it straightforward for users to navigate through various options. This simplicity is particularly beneficial for new users who may find CPAP therapy overwhelming.

In addition to its advanced technological features, the S8 Series II incorporates a compact and lightweight design, making it highly portable. Users can easily transport the device for travel or overnight stays, promoting compliance with their therapy even when away from home.

Moreover, the S8 AutoScore II comes equipped with a SmartStart function, which allows the device to automatically commence therapy once the user breathes into the mask. This feature enhances user convenience, eliminating the need for manual adjustments once the device is in place.

The S8 Series II also emphasizes quiet operation, ensuring that the device does not disturb the user or their partner during the night. Low noise levels and a comfortable fit contribute to an overall positive sleeping experience, which is crucial for the success of CPAP therapy.

Overall, the ResMed S8 Series II, particularly the S8 AutoScore II model, represents a blend of cutting-edge technology, user-centric design, and versatile functionality. These characteristics make it a reliable choice for individuals seeking effective management of obstructive sleep apnea while prioritizing comfort and convenience in their treatment journey.