9How to Use the Menus

English

The below example shows the progression of screens as you navigate to the Usage Data

from the Patient Summary Menu.

The data displayed in the Summary menu screens is view only. In some Summary menu

screens, pressing Enter when the soft key in the title bar is selected allows you to choose

a time period over which the data is measured (eg, a day, a week, a month, etc.).

Patient Settings Menu

The Patient Settings menu allows you to view and change settings such as mask type, tube

length and the humidifier used. You can also access the Mask-Fit feature.

!
WARNING

If these settings do not match your system set-up, this may alter the pressure

you actually receive and reduce the effectiveness of your treatment.

123
SUMMARY 21:53
EFFICACY DATA
USAGE DATA
ALARM HISTORY
EVENT HISTORY
SETTINGS SUMMARY
SERVICING
LK: 1.7 Vt: 780 RR: 10 MV: 7.8
SUMMARY 21:53
EFFICACY DATA
USAGE DATA
ALARM HISTORY
EVENT HISTORY
SETTINGS SUMMARY
SERVICING
LK: 1.7 Vt: 780 RR: 10 MV: 7.8
21:53
SUMMARY
6mths
USAGE DATA
Used Since:
12 Oct 2006
Used:
740
hrs
102/106
days
Usage: 7:15
hrs/day
LK: 1.7 Vt: 780 RR: 10 MV: 7.8
SETTINGS
21:53
LK:
1.7
Vt:
780
RR:
10
MV:
7.8
Mask: MIR FULL
Humidifier: None
Tube Length: 2m
Check Mask Fit
SmrtStrt/Stp: Off
Function Default Function Description Settings
Mask MIR
FULL
Selects your mask type. See www.resmed.com or
ask your clinician for the
correct setting for your mask
type.
Humidifier NONE Selects the type of humidifier to
be used with the device.
NONE, H2i (HumidAire 2iC),
PASSOVER, HUMIDAIRE
If the HumidAire 2i is used, it
is automatically detected
and H2i is displayed.
Tube Length 2 m Selects the length of air tubing
connecting your mask to the
device.
2 m,
3 m
Check Mask-
Fit
Allows you to check your mask-
fit star rating.
View only
248603 VPAP III ST-A with QuickNav User Guide EUR1.book Page 9 Tuesday, January 17, 2012 11:25 AM