6.MAINTENANCE AND CARE
WARNING
Before entry into the instrument compartment, refrigeration compartment, or treatment
chamber ISOLATE THE CABINET FROM MAINS POWER unless otherwise stated.
The cabinet has been designed and built for a long life and requires minimal attention
and maintenance. However, regular attention to the few points will ensure a long and
trouble free life.
If the cabinet does fail, the expertise of our service engineers is readily available for
either diagnostic advice or on-site attendance. Service contracts are available after the
one year guarantee period finishes and details are available from our Service
Department.
This section contains a maintenance schedule to follow and a methods section giving
instructions relating to the schedule.
Note: If the cabinet is going to be left unused for more than a month, we suggest that
the cabinet is started up and run through all its functions on a monthly basis. This will
ensure that the cabinet is in perfect running order for when it is required.

6.1 MAINTENANCE SCHEDULE

Weekly or between tests
Clean the cabinet's interior and exterior surfaces
Clean drain(s)
Clean atomiser(s)
Check wicks
Month
Renew wicks
Lubricate moving parts
Check safety thermostat
Clean refrigerator condenser
Check distilled water filter
19