Sanyo 550 instruction manual

Models: 550

1 73
Download 73 pages 5.1 Kb
Page 57
Image 57

3 bytes: Relative Humidity in %, eg "095"

starts byte No 012

 

ends

byte No 014

6 bytes: Temperature Setpoint in degC, eg "-150.0"

starts

byte No 015

 

ends

byte No 020

3 bytes: Relative Humidity Setpoint in %, eg "095"

starts byte No 021

 

ends

byte No 023

3 bytes: Third Variable setpoint in %, eg "050"

starts

byte No 024

 

ends

byte No 026

3 bytes: Intelligence Table Line, eg "001"

starts byte No 027

 

ends

byte No 029

7 bytes: Temperature P term, eg "-123.45

starts byte No 030

 

ends

byte No 036

7 bytes: Temperature I term, eg "-123.45"

starts byte No 037

 

ends

byte No 043

7 bytes: Temperature D term, eg "-123.45"

starts byte No 044

 

ends

byte No 050

3 bytes: Heating Effort in %, eg "100"

starts byte No 051

 

ends

byte No 053

3 bytes: Cooling Effort in %, eg "100"

starts

byte No 054

 

ends

byte No 056

7 bytes: Relative Humidity P term, eg "-123.45"

starts byte No 057

 

ends

byte No 063

7 bytes: Relative Humidity I term, eg "-123.45"

starts

byte No 064

 

ends

byte No 070

7 bytes: Relative Humidity D term, eg "-123.45

starts byte No 071

 

ends

byte No 077

3 bytes: Humidification Effort in %, eg "100"

starts

byte No 078

 

ends

byte No 080

3 bytes: Dehumidification Effort in %, eg "100"

starts

byte No 081

 

ends

byte No 083

8 bytes: Running Digital Channels, eg "10110010"

starts

byte No 084

 

ends

byte No 091

6 bytes: Temperature Setpoint No 1 in degC, eg "-150.0" (f500 only) starts byte No 092

ends byte No 097

3 bytes: Relative Humidity Setpoint No 1 in %,

eg "050" (f500 only)starts byte No 098 ends byte No 100

6 bytes: Temperature Setpoint No 2 in degC, eg "-150.0" (f500 only) starts byte No 101 ends byte No 106