Instructions for the installer
2. INSTALLATION OF THE STERILIZER
The sterilizer must be set up on a perfectly flat surface and left standing free. Make sure to allow enough room at the back for easy access to the
terminal strip for connection of the power cable to mains.
Installation must be carried out by a qualified technician in compliance with applicable regulations in force.
2.1 Electrical connection
Make sure that the power line voltage matches the specifications indicated on the rating plate located to the rear of the sterilizer. This rating plate must never be removed.
The sterilizer is supplied complete with a
The standard cable is of the
If necessary the electrical supply cable should be replaced with an
The (yellow/green) earth cable on the end of the cable to be connected to the sterilizer should be at least 20 mm longer.
Make sure that plug and wall socket are of the same type and are in conformity with regulations in force. Also check that the power supply line is provided with adequate earthing. Avoid using adapters or shunters.
In case of permanent connection, ensure that power cable is equipped with an easily accessible omnipolar switch fitted in proximity to the apparatus and with a gap between contacts equal to or greater than 3 mm.