CHARGING THE ACCUMULATOR

ACCUMULATOR TESTING

PROCEDURE

To check or charge the accumulator the following equip- ment is required:

31254 Charge Kit: which includes the following: (Shown in Figure 2 on the next page.)

Accumulator Tester (Part Number 02835).

Charging Assembly (P/N 15304) Includes a liquid filled gauge w/snub valve, hose and charge fitting.

NITROGEN bottle with a 800 psi/56 bar minimum charge. (Not included in 31254 Charge Kit.)

1.Remove the valve cap assembly from the sinker drill.

2.Remove the protective cap and loosen the 5/8-inch hex locking nut on the tool charging valve 1-1/2 turns.

3.Holding the chuck end of Accumulator Tester (P/N 02835) turn the gauge fully counterclockwise to en- sure that the stem inside the chuck is completely retracted.

4.Thread the tester onto the accumulator charg- ing valve. Do not advance the gauge-end into the chuck-end. Turn as a unit. Seat the chuck on the accumulator charging valve and hand tighten only.

5.Advance the valve stem of the tester by turning the gauge-end clockwise until a pressure is read on the gauge (charge pressure should be 600-700 psi/42- 48 bar).

6.If pressure is OK unscrew the gauge-end from the chuck to retract the stem, then unscrew the entire tester assembly from the accumulator charging valve. If pressure is low, charge the accumulator as described in the following procedure.

7.Tighten the 5/8-inch hex locking nut on the tool charging valve. Be careful not to overtighten. Install the protective cap and valve cap assembly.

ACCUMULATOR CHARGING

1.Perform steps 1 through 4 of the accumulator test- ing procedure above.

2.Connect the chuck of the charging assembly to the charging valve on the accumulator tester or, if pre- ferred, remove the tester from the charging valve and connect the charging assembly chuck directly to the charging valve.

3.Adjust the regulator to the charging pressure of 600 psi/42 bar.

NOTE:

It may be necessary to set the gauge at 650-700 psi/45-48 bar to overcome any pressure drop through the charging system.

4.Open the valve on the charging assembly hose.

5.When the accumulator is fully charged close the valve on the charging assembly hose and remove the charging assembly chuck from the accumulator tester or tool charging valve.

6.If the accumulator tester has been used, be sure to turn the gauge-end fully counterclockwise before re- moving the tester from the charging valve of the tool.

7.Tighten the 5/8-inch hex locking nut on the tool charging valve and replace the protective cap.

8.Replace the valve cap assembly.

GENERAL SERVICE NOTES

1.If the breaker is repainted after servicing, be sure to mask off the vent in the valve cap assembly. Do not allow paint to enter the IN and OUT ports or the bore of the foot assembly.

2.If the handle grips need to be replaced.

a.Remove the old grips and clean the handle.

b.Wash the new grips and the handle clean and dry, simply push or drive the grips on. DO NOT lubricate the parts. The grips will not be secure on the handle if any grease or oil is used.

14 ► SK58 User Manual

Page 14
Image 14
Stanley Black & Decker SK58 user manual Charging the Accumulator, Accumulator Testing Procedure, Accumulator Charging