8

E n g l i s h

 

 

 

 

 

 

 

 

 

 

V

.

 

E

M

I

 

(

E

l

e

c

t

r

o

m

a

g

n

e

t

i

c

 

I

n

t

e

r

f

e

r

e

n c e )

 

 

 

D. DISTANCE FROM THE SOURCE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EM energy rapidly becomes more intense as you get closer to the source. For this reason, EMI from hand-held devices is of special concern. (See C.1) A person using one of these devices can bring high levels of EM energy very close to your chair without you knowing it.

E.IMMUNITY LEVEL

1.The level of EM energy is measured in volts per meter (V/m). Every power wheelchair can resist EMI up to a certain level. This is called its “immunity level”.

2.The higher the immunity level, the less the risk of EMI. It is believed that a

20 V/m immunity level will protect the power wheelchair user from the more common sources of radio waves.

3.The configurations tested and found to be immune to at least 20 V/m are:

a.Quickie Aspire M11 power wheelchair with a right-hand mounted Penny & Giles VSI integral controller, highback seat, flip-back armrests, solid flip-up footplate and U1 gel cell batteries.

b.Quickie Aspire F11 power wheelchair with a right-hand mounted Penny & Giles VSI integral controller, highback seat, flip-back armrests, solid flip-up footplate and Gp24 gel cell batteries.

Individuals with physical limitations requiring the use of a specialty control input device known not to be immune to 20V/m, or not known, should exercise extra care around known sources of EMI.

There is no way to know the effect on EMI if you add accessories or modify this chair. Any change to your chair may increase the risk of EMI. Parts from other suppliers have unknown EMI properties.

F. REPORT ALL SUSPECTED EMI INCIDENTS

You should promptly report any unintended movement or brake release. Be sure to indicate whether there was a radio wave source near your chair at the time. Contact: Sunrise Medical Customer Service Department at (800) 333-4000.

930756 Rev. B

Page 8
Image 8
Sunrise Medical M11, F11 instruction manual Immunity Level, Report ALL Suspected EMI Incidents