Service Information
30 P6434 Probe Instructions
Functional Verification
This procedure checks the basic functionality of the probe by verifying that the
probes recognize signal activity at the probe tips. No calibration is necessary.
The functional verification procedure for the P6434 probe requires the following
test equipment:
HAdjustment and verification test fixture, part number 671-3599-00
HPower supply for the test fixture (refer to the Optional Accessories list in the
Replaceable Parts section for part number information)
HLogic analyzer mainframe
HLogic analyzer module
To perform the functional verification procedure, follow these steps:
1. Connect the module ends of the P6434 probe to the logic analyzer module.
2. Connect the probe tip to J5 on the the test fixture as shown in Figure 25.
J5
J15
Figure 25: Jumper and connector locations on the test fixture
3. Place J15 in the INT position to select the internal 50 MHz clock.
4. Connect the power adapter to the test fixture and power on the test fixture.
5. Power on the logic analyzer.
6. Verify that the logic analyzer passes the power-on diagnostics.