EDIT Menu

Figure 3 ￿78: Count Up Pattern Display

The following parameters can be set:

Step

The number of points for each data item. This value can be

 

set to 1 ± 10.

Max

The maximum value for the data. This value can be set

 

anywhere between (Min value + 1) and 4095.

Min

The minimum value for the data. This value can be set

 

anywhere between 0 and (Max value ± 1).

￿Count Down. Data is created sequentially, decreasing from the initial value designated for Max to the minimum value designated for Min. When the minimum value is exceeded, the value returns to the maximum value and countdown begins again. The values for Step, Max and Min can be set in the same manner as with Count Up.

￿Gray Code. A data pattern is created in which the difference between adjacent data is only 1 bit. When the 4096 items of 12-bit data end, the data repeats from the beginning. In the example shown in Figure 3 -79, the waveform point size has been set to 4096 and Step has been set to 1. Only the Step parameter is set for the Gray Code pattern option.

3 ￿102

AWG2021 User Manual