SETUP Menu

To set the Source to External:

6.Press the Source button in the side menu and select External. The clock icon shown below will be displayed on the screen.

The CH1 and CH2 clocks are controlled with the external clock input from the CLOCK IN connector on the rear panel. The external clock frequency that can be input to the connector is up to 250 MHz.

CH1 Waveform Operations

CH1 Operation is used to operate the CH1 waveform and the waveform from CH2, or the CH1 waveform and a waveform input from an external source. It then outputs the result from the CH1 output connector. Two types of operation

Ðaddition (Add) or multiplication (AM) Ð can be performed for the CH1 waveform and CH2 waveform or external signal.

To set the desired operation process for the CH1 waveform:

1.Select CH1 Operation from the bottom menu.

2.Select the desired process from the side menu. The following choices are available:

Standard Type

Option 02

 

 

Normal

Normal

External AM

AM

 

Add

 

External AM

 

 

The following is a more detailed description of each process:

Normal

Waveforms are output separately from each channel.

AM

CH1 output is the CH1 waveform with the amplitude

 

modulated (multiplied) by the waveform output from CH2.

 

See Figure 3 -156.

AWG2021 User Manual