Appendix C: Performance Verification
Equipment | None |
Required |
|
|
|
Prerequisites | Power on the AW2021 and allow a 20 minute warmup period before |
| doing this procedure. |
|
|
1.Verify that internal diagnostics pass: Do the following substeps to verify passing internal diagnostics.
a.Display the diagnostics menu and select all tests: Push UTILITYDiag/CalDiagnostics xxxxAll.
The list on the left shows the tests available for diagnostics. In addition to selecting all of the tests shown for diagnostics, you can select only the test you want to run using the general purpose knob. In Figure
Figure C1: Diagnostics Menu
b.Run the diagnostics: Select Execute from the side menu. This executes all the AWG2021 diagnostics automatically.
c.Wait: The internal diagnostics do an extensive verification of AWG2021 functions. While it progresses, the screen displays the busy icon. When finished, the resulting status appears on the screen.
d.Confirm that no failures are found: Verify that no failures are found and reported
2.Return to regular service: Push a button (other than UTILITY) in the MENU column to exit the diagnostic menu.
AWG2021 User Manual | C3 |