Appendix C: Performance Verification

Equipment

None

Required

 

 

 

Prerequisites

Power on the AW￿2021 and allow a 20 minute warmup period before

 

doing this procedure.

 

 

1.Verify that internal diagnostics pass: Do the following substeps to verify passing internal diagnostics.

a.Display the diagnostics menu and select all tests: Push UTILITY￿Diag/Cal￿Diagnostics xxxx￿All.

The list on the left shows the tests available for diagnostics. In addition to selecting all of the tests shown for diagnostics, you can select only the test you want to run using the general purpose knob. In Figure C-1, the symbol (*) to the left of Diagnostics column indicates the tests selected.

Figure C￿1: Diagnostics Menu

b.Run the diagnostics: Select Execute from the side menu. This executes all the AWG2021 diagnostics automatically.

c.Wait: The internal diagnostics do an extensive verification of AWG2021 functions. While it progresses, the screen displays the busy icon. When finished, the resulting status appears on the screen.

d.Confirm that no failures are found: Verify that no failures are found and reported on-screen. If the diagnostics displays Fail as the result of any test, contact your nearest representative.

2.Return to regular service: Push a button (other than UTILITY) in the MENU column to exit the diagnostic menu.