Appendix C: Performance Verification
Calibration The AWG2021 includes internal calibration routines that check electrical characteristics such as amplitude, offset, trigger level, clock, filters, X5 output amplifier and attenuation, and adjust internal calibration constants as necessary.
Equipment Required | None |
|
|
Prerequisites | Power on the AW2021 and allow a 20 minute warmup period |
| before doing this procedure. |
|
|
1.Verify that internal adjustments pass: Do the following substeps to verify passing of internal adjustments.
a.Display the calibrations menu and select all tests: Push UTILITYDiag/CalCalibrations xxxxAll. See the menu in Figure
Figure C2: Calibrations Menu
b.Run the calibrations routine: Select Execute from the side menu. This executes the AWG2021 calibrations routine automatically.
c.Wait: The internal calibration does an exhaustive verification of proper AWG2021 function. While it progresses, the busy icon appears on screen. When finished, the resulting status will appear on the screen.
d.Confirm that no failures are found: Verify that no failures are found and reported
2.Return to regular service: Push any button (other than UTILITY) in the MENU column to exit the diagnostic menu.