Appendix C: Performance Verification
Table C,1: Test Equipment (Cont.)
Item | Minimum Requirements | Example | Purpose |
|
|
|
|
Digital multimeter | DC volts range: 0.05:V to 5:V | Fluke 8842A | Used throughout the checks |
| Accuracy: 0.1% |
| to measure voltage. |
|
|
|
|
Function generator | Output voltage: | Tektronix FG5010 Program; | Used to input the trigger |
|
| mable Function Generator* | signal. |
Power supply (Option 03) | Output voltage: | Tektronix PS5010 Program; | Used to check ECL digital |
|
| mable Power Supply* | data output. |
Performance Check/Adjust; | Must use example listed | Tektronix Part 063;2171;XX | Used throughout the checks |
ment disk |
|
| to provide waveform files. |
|
|
|
|
*Requires a TM 5000 Series Power Module Mainframe.
Performance Check/ Adjustment Files
Table
NOTE. The files on the Performance Check/Adjustment disk are locked (the files names are displayed with *), so the data in these files cannot be changed unless the lock is opened. The file data includes not only waveform data, but also output parameters.
When you select a file with the Waveform Sequence item, the AWG2021 output parameters change to those specified in the file, and the waveform output reflects waveform data in the file. After selecting a file, do not change an output parameter with the SETUP menu unless a procedure instructs you to do so. During the procedures, if you are unsure that the AWG2021 settings still match the file's settings, select the waveform again using the Waveform Sequence item on the SETUP menu.
HF_LF.WFM file on the Performance Check/Adjustment disk is used to adjust the AWG2021 externally. On adjustment, see the AWG2021 Service Manual.