Appendix C: Performance Verification

Table C,1: Test Equipment (Cont.)

Item

Minimum Requirements

Example

Purpose

 

 

 

 

Digital multimeter

DC volts range: 0.05:V to 5:V

Fluke 8842A

Used throughout the checks

 

Accuracy: ￿0.1%

 

to measure voltage.

 

 

 

 

Function generator

Output voltage: -5:V to 5:V

Tektronix FG￿5010 Program;

Used to input the trigger

 

 

mable Function Generator*

signal.

Power supply (Option 03)

Output voltage: -2:V

Tektronix PS￿5010 Program;

Used to check ECL digital

 

 

mable Power Supply*

data output.

Performance Check/Adjust;

Must use example listed

Tektronix Part 063;2171;XX

Used throughout the checks

ment disk

 

 

to provide waveform files.

 

 

 

 

*Requires a TM 5000 Series Power Module Mainframe.

Performance Check/ Adjustment Files

Table C-2 lists the waveform files on the Performance Check/Adjustment disk (063-2171-XX) that are used in these performance tests, the AWG2021 front-panel settings that each file sets up, and the performance test that uses each file.

NOTE. The files on the Performance Check/Adjustment disk are locked (the files names are displayed with *), so the data in these files cannot be changed unless the lock is opened. The file data includes not only waveform data, but also output parameters.

When you select a file with the Waveform Sequence item, the AWG2021 output parameters change to those specified in the file, and the waveform output reflects waveform data in the file. After selecting a file, do not change an output parameter with the SETUP menu unless a procedure instructs you to do so. During the procedures, if you are unsure that the AWG2021 settings still match the file's settings, select the waveform again using the Waveform Sequence item on the SETUP menu.

HF_LF.WFM file on the Performance Check/Adjustment disk is used to adjust the AWG2021 externally. On adjustment, see the AWG2021 Service Manual.

AWG2021 User Manual

C,7