Appendix C: Performance Verification

2.Set the AWG2021 controls and select the waveform file:

a.Initialize AWG2021 controls: Push UTILITY￿Misc￿Config...￿Re- set to Factory￿O.K.

b.Select the file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Turn the general purpose knob to display the list of waveform files and highlight the MODE.WFM file.

￿Push ENTER to select the file. This button is located to the lower-right of the numeric keypad.

3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check cont mode: Check that the amplitude of the sine wave displayed on the oscilloscope is 5 vertical divisions and that 5 cycles of the waveform are displayed.

5.If Option 02 is installed (Option 02 adds the CH2 output channel): Repeat this procedure, connecting the oscilloscope to the AWG2021 CH2 output connector, pushing Waveform Sequence to select the waveform for CH2, and turning on the CH2 output.

6.End procedure: Disconnect the oscilloscope.

Check Triggered Mode

Electrical

￿pe￿a ing mode, T￿igge￿ed, on page$B%1.

Characteristic Checked

 

 

 

Equipment Required

Two 50$W coaxial cable￿, a f!nc ion gene￿a o￿, and an

 

o￿cillo￿cope.

 

 

Prerequisites

The AWG2021 mee ￿ he p￿e￿eq!i￿i e￿ li￿ ed on page C%5.

 

 

1.Install the test hookup and set test equipment controls:

a.Hook up the oscilloscope: Connect the AWG2021 CH1 output connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope.

b.Hook up the function generator:

￿Connect the AWG2021 TRIGGER INPUT connector through a coaxial cable to the function generator output connector (see Figure C-4).

AWG2021 User Manual

C￿11