Appendix C: Performance Verification
2.Set the AWG2021 controls and select the waveform file:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Select the file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to display the list of waveform files and highlight the MODE.WFM file.
Push ENTER to select the file. This button is located to the
3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check cont mode: Check that the amplitude of the sine wave displayed on the oscilloscope is 5 vertical divisions and that 5 cycles of the waveform are displayed.
5.If Option 02 is installed (Option 02 adds the CH2 output channel): Repeat this procedure, connecting the oscilloscope to the AWG2021 CH2 output connector, pushing Waveform Sequence to select the waveform for CH2, and turning on the CH2 output.
6.End procedure: Disconnect the oscilloscope.
Check Triggered Mode
Electrical | pea ing mode, Tiggeed, on page$B%1. |
Characteristic Checked |
|
|
|
Equipment Required | Two 50$W coaxial cable, a f!nc ion genea o, and an |
| ocillocope. |
|
|
Prerequisites | The AWG2021 mee he peeq!ii e li ed on page C%5. |
|
|
1.Install the test hookup and set test equipment controls:
a.Hook up the oscilloscope: Connect the AWG2021 CH1 output connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope.
b.Hook up the function generator:
Connect the AWG2021 TRIGGER INPUT connector through a coaxial cable to the function generator output connector (see Figure
AWG2021 User Manual | C11 |