Appendix C: Performance Verification
2.Set the AWG2021 controls and select the waveform file:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Modify the AWG2021 default settings:
Push MODEGatedPolarity to highlight Positive.
Select Impedance from the side menu to highlight 1 MW.
c.Select the file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE.WFM file, using the general purpose knob.
Push ENTER to select the file.
3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check gated mode with manual trigger: Push and hold the AWG2021 MANUAL TRIGGER button, and check that the oscilloscope continuously displays a sine wave while the MANUAL TRIGGER button is pushed.
5.Check gated mode with gate signal:
a.Apply gate signal: Turn function generator output on.
b.Check gated mode with positive gate signal: Check that the oscilloscope displays a sinewave when the gate signal is high.
AWG2021 User Manual | C17 |