Appendix C: Performance Verification

2.Set the AWG2021 controls and select the waveform file:

a.Initialize AWG2021 controls: Push UTILITY￿Misc￿Config...￿Re- set to Factory￿O.K.

b.Modify the AWG2021 default settings:

￿Push MODE￿Gated￿Polarity to highlight Positive.

￿Select Impedance from the side menu to highlight 1 MW.

c.Select the file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Highlight the MODE.WFM file, using the general purpose knob.

￿Push ENTER to select the file.

3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check gated mode with manual trigger: Push and hold the AWG2021 MANUAL TRIGGER button, and check that the oscilloscope continuously displays a sine wave while the MANUAL TRIGGER button is pushed.

5.Check gated mode with gate signal:

a.Apply gate signal: Turn function generator output on.

b.Check gated mode with positive gate signal: Check that the oscilloscope displays a sinewave when the gate signal is high.

AWG2021 User Manual

C￿17