Appendix C: Performance Verification

￿Push MODE￿Waveform Advance￿Slope to highlight Positive.

￿Select Level from the side menu, and turn the general purpose knob to select a 1.0 V level.

￿Select Impedance from the side menu to highlight 50 W. c. Select waveform file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Highlight the MODE_ADV.SEQ file using the general purpose knob.

￿Push ENTER to select the file.

3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check waveform advance: Repeatedly push the AWG2021 MANUAL TRIGGER button, and check that the oscilloscope displays a continuous sine wave that switches between two frequencies at each manual trigger.

5.End procedure: Disconnect the oscilloscope.

Check Autostep Mode

Electrical

￿perating mode, Autostep, on page#B$1.

Characteristic Checked

 

 

 

Equipment Required

Two 50#W coaxial cables and an oscilloscope.

 

 

Prerequisites

The AWG2021 meets the prerequisites listed on page C$5.

 

 

1.Install test hookup and set test equipment controls: a. Hook up the oscilloscope:

￿Connect the AWG2021 CH1 output through a coaxial cable to the oscilloscope CH1 vertical input.

￿Connect the AWG2021 CH1 SYNC Out output through a coaxial cable to the oscilloscope CH1 vertical input (see Figure C-10).

C￿20

AWG2021 User Manual