Appendix C: Performance Verification
Push MODEWaveform AdvanceSlope to highlight Positive.
Select Level from the side menu, and turn the general purpose knob to select a 1.0 V level.
Select Impedance from the side menu to highlight 50 W. c. Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE_ADV.SEQ file using the general purpose knob.
Push ENTER to select the file.
3.Turn on the AWG2021 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check waveform advance: Repeatedly push the AWG2021 MANUAL TRIGGER button, and check that the oscilloscope displays a continuous sine wave that switches between two frequencies at each manual trigger.
5.End procedure: Disconnect the oscilloscope.
Check Autostep Mode
Electrical | perating mode, Autostep, on page#B$1. |
Characteristic Checked |
|
|
|
Equipment Required | Two 50#W coaxial cables and an oscilloscope. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C$5. |
|
|
1.Install test hookup and set test equipment controls: a. Hook up the oscilloscope:
Connect the AWG2021 CH1 output through a coaxial cable to the oscilloscope CH1 vertical input.
Connect the AWG2021 CH1 SYNC Out output through a coaxial cable to the oscilloscope CH1 vertical input (see Figure
C20 | AWG2021 User Manual |